How to Measure Microbial Growth Using a Hemocytometer

By Dr. Zubair Khalid, DVM, MS, PhD ·

How to Measure Microbial Growth Using a Hemocytometer

A hemocytometer turns an ordinary microscope into a volumetric counting instrument. It gives a direct count of cells per milliliter in a chamber of known depth, and that number is the raw material for growth curves, doubling-time calculations, and inoculum standardization. Optical density reads turbidity and impedance reads metabolic activity, but the hemocytometer counts physical particles. That distinction matters when you need absolute cell numbers or when you need to separate live from dead cells with a viability dye [1][2].

Use the hemocytometer when the organism is large enough to resolve under light microscopy, such as yeasts, spores, algae, protozoa, and the larger bacilli. Use it when a plate count would cost 24 to 48 hours or when you want an independent cross-check of an OD-based growth curve. Do not use it as a stand-in for a viable count of small rods unless you pair it with a viability stain and a phase-contrast or darkfield microscope.

Time budget: 5 to 10 minutes for sample preparation and dilution, 3 to 5 minutes for chamber filling and settling, 5 to 10 minutes to count one chamber, and about 2 minutes for the arithmetic. A duplicate count on both sides of the slide takes 25 to 40 minutes of hands-on time. A full growth curve with eight timepoints and duplicate counts runs 5 to 7 hours of elapsed bench time, most of it incubation. The measurement itself is quick. The discipline is in the mixing, the dilution, and the counting rules.

What the Hemocytometer Actually Measures

Chamber geometry and the counting volume

A hemocytometer is a thick glass slide with a polished central platform and a ruled grid etched into the glass. A coverslip rests on raised supports, leaving a gap of fixed depth between the coverslip and the grid. The most common format, the improved Neubauer chamber, has a depth of 0.1 mm (100 µm). The central counting area is 1 mm by 1 mm and is subdivided into 25 medium squares, each 0.2 mm by 0.2 mm. Each medium square is further divided into 16 small squares.

Those dimensions define the volume. A single 1 mm by 1 mm square at 0.1 mm depth holds 0.1 mm³ of liquid. Since 1 mm³ equals 1 µL, that square holds 0.1 µL, or 10⁻⁴ mL. Multiplying a count by 10,000 converts cells per square to cells per milliliter. That is where the familiar 10,000 multiplier comes from.

Different rulings change the multiplier. If you count only the five medium squares (four corners plus center) inside the central 1 mm² area, the total area is 0.2 mm² and the volume is 2 × 10⁻⁵ mL. The correct multiplier per counted cell is then 50,000, not 10,000. The multiplier is a property of the area you count, not of the slide in general. Getting this wrong is one of the most frequent arithmetic errors in hemocytometer work.

For bacteria, the standard chamber is the Petroff-Hausser counting chamber, which has a depth of 0.02 mm. That shallower depth reduces the focal range and improves resolution for cells around 1 µm. Its multiplier for the full 1 mm² area is 50,000.

Total count versus viable count

The hemocytometer counts particles. A dead yeast cell, a live yeast cell, a bud, and a chunk of debris of the right size all register the same way. That is a total count, and it is the correct tool when you want biomass or particle number but the wrong tool when you want colony-forming units.

Plate counting measures colony-forming units, or CFU, which are viable cells capable of dividing on agar. The CFU count is a viable count but it is not a complete viable count. Sublethally injured cells may fail to form colonies even though they remain metabolically active, and a double subculture method has been described specifically to enumerate that injured fraction by comparing solid-medium and liquid-medium recovery [3]. Viable but non-culturable cells are invisible to plate counting entirely.

The practical consequence is that a hemocytometer count is usually higher than a CFU count from the same sample, sometimes by a factor of two to ten. That gap does not mean one method is wrong. It means the two methods answer different questions. A hemocytometer tells you how many cells are present. A plate count tells you how many can grow into colonies under the chosen conditions.

When the hemocytometer is the right choice

The hemocytometer is the right choice in four situations. First, when you need an answer within an hour and a plate count would take a day. Second, when you are preparing an inoculum and need to standardize cell density before an experiment. Third, when the organism does not form reliable colonies on agar, as with many protozoa and algae. Fourth, when you want to cross-check an OD reading that may have drifted because of pigmentation, clumping, or dead cells.

It is the wrong choice when the cells are smaller than about 1 µm and you lack phase contrast, when the sample is full of debris of the same size as the cells, or when you need a viable count and cannot use a dye.

Materials and Reagents

ItemSpecification or working concentrationPurpose
Hemocytometer slideImproved Neubauer, 0.1 mm depthCounting yeasts, spores, algae, protozoa, large bacteria
Petroff-Hausser chamber0.02 mm depthCounting small bacteria
Coverslip22 mm by 22 mm, #1.5 thicknessDefines the chamber volume
Micropipettes10 µL and 100 to 1000 µL, sterile tipsCharging the chamber and making dilutions
Phosphate-buffered saline10 mM, pH 7.2 to 7.4, sterileDiluent that preserves cell shape
Trypan blue stock0.4% in PBS, working concentration 0.2% after 1:1 mixingViability dye for yeast and mammalian cells
Methylene blue0.1% working concentrationViability dye for yeast
Formalin10% neutral buffered, added to 2 to 5% finalFixation before transport out of the cabinet
Ethanol70%Cleaning the slide and coverslip
Lens paperLint-freeDrying without scratching the ruling
Phase-contrast microscope10x and 40x objectives, optional 100x oilResolving cells against the grid
Hand tally counterMechanical or electronicCounting without losing track
Vortex mixerVariable speedResuspending settled cells

Working concentrations matter. Trypan blue is sold as a 0.4% stock solution. Mixing it 1:1 with cell suspension gives a 0.2% working concentration, which is the standard for viability counting. Stronger dye increases background staining and makes live cells look dead. Methylene blue for yeast is used at 0.1% and read within a few minutes, because yeast cells slowly reduce the dye and lose the stain.

Safety note

Clinical and environmental samples may contain infectious agents. Handle them in a biosafety cabinet at the appropriate containment level, wear gloves and a lab coat, and disinfect the slide, the pipette tips, and the bench with a validated disinfectant after counting. Trypan blue is a known teratogen in some models and stains skin and clothing. Wear gloves and eye protection, and avoid generating aerosols. Formalin is a respiratory irritant and a carcinogen, so fix samples in a fume hood or a ventilated cabinet. Glass hemocytometers break, and broken coverslips are sharp.

Sample Preparation and Dilution

Hitting the counting window

The count is only as good as the number of cells in the field. The target window is 100 to 200 cells total across the squares you plan to count, which corresponds to roughly 20 to 50 cells per 1 mm² square when you count five squares. Below 20 cells per square, the Poisson sampling error becomes large. Above 100 cells per square, cells overlap, stick to the lines, and get missed. Both extremes produce a count that looks precise but is not.

The dilution calculation uses the standard relation C₁V₁ = C₂V₂. If you expect 1 × 10⁷ cells per mL and you want 5 × 10⁵ cells per mL in the chamber, you need a 20-fold dilution. Prepare it in two steps (a 1:5 followed by a 1:4) rather than one large step, because small volume errors are magnified in a single large dilution.

Record the dilution factor explicitly. When a viability dye is mixed 1:1 with sample, the dye contributes a factor of 2 on top of any PBS dilution. A sample diluted 1:10 in PBS and then 1:1 in trypan blue has a total dilution factor of 20.

Mixing and disaggregation

Microbial cells settle quickly, especially yeasts and spores. Vortex for 10 to 15 seconds or invert the tube 10 times immediately before drawing the sample. Do not vortex fragile protozoa or filamentous algae at high speed, because shear damages them. For clumped cultures, a brief sonication in a water bath or the addition of a small amount of EDTA helps, provided the organism tolerates it. Do not add detergent to a viability count, because it lyses cells.

Fixation and staining

Fixation is optional for a quick count. It is useful when the organism is motile, when the sample must leave the cabinet, or when counting will be delayed. Add neutral buffered formalin to a final concentration of 2 to 5%, or add ethanol to the same final range. Fixed cells shrink slightly, which biases the count downward by a few percent. Run a paired unfixed and fixed count once for each new organism to quantify that bias.

For viability, mix the sample with trypan blue at 1:1 and count within 5 minutes. Live cells exclude the dye and stay clear. Dead cells lose membrane integrity, take up the dye, and appear blue. For yeast, methylene blue at 0.1% gives a similar readout. Count the total cells and the unstained cells separately, then calculate the viability fraction as unstained divided by total.

How to Measure Microbial Growth Using a Hemocytometer Step by Step

The procedure below assumes an improved Neubauer chamber and a yeast or spore suspension. Adapt the counting squares and the multiplier for other chambers as described in the variations section.

  1. Clean the slide and coverslip. Wipe both with 70% ethanol and dry with lens paper. Residual protein or fingerprints create uneven wetting and bubbles. Why: a dirty chamber fills irregularly and distorts the focal plane.
  1. Seat the coverslip. Place the coverslip on the chamber supports and press gently with a gloved finger until Newton's rings (interference fringes) appear at the edges. Why: a loose coverslip increases the effective depth and inflates the volume, which deflates the final count.
  1. Mix the sample. Vortex for 10 to 15 seconds or invert 10 times. Why: settled cells produce counts that are too low in the first chamber and too high in the last.
  1. Prepare the dilution. Dilute in sterile PBS to the target window and record the dilution factor. Why: counting outside the 20 to 100 cells per square window adds error that no arithmetic can remove.
  1. Add the viability dye if needed. Mix sample and 0.4% trypan blue 1:1, wait 1 to 2 minutes, and count within 5 minutes. Why: dye uptake by dead cells is fast, but prolonged exposure also stains stressed live cells.
  1. Draw 10 µL. Use a micropipette with a fresh tip. Why: a calibrated pipette delivers a reproducible volume and avoids the air bubbles that a Pasteur pipette introduces.
  1. Charge the chamber. Touch the tip to the V-shaped notch at the edge of the coverslip and let capillary action pull the liquid in. Stop when the chamber is full and the liquid reaches the far edge of the coverslip but not the moat. Why: capillary filling avoids shear and bubbles, and overfilling floods the moat and changes the depth.
  1. Fill the second chamber if present. Repeat with a fresh 10 µL aliquot. Why: duplicate chambers detect pipetting and distribution errors that a single count cannot.
  1. Let cells settle for 2 to 3 minutes. Place the slide on a level bench, not on a warm microscope stage. Why: cells in different focal planes are missed, and warm stages accelerate evaporation.
  1. Locate the grid at 10x. Focus on the ruled lines first, then move to 40x and fine-focus on the cells. Why: focusing on the cells first often puts the grid out of focus and makes the boundaries ambiguous.
  1. Choose the counting squares. For the improved Neubauer, count the four corner medium squares plus the center medium square (five squares), or count the entire central 1 mm² area. Why: the multiplier depends on the area counted, so decide before you start.
  1. Apply the boundary rule. Count cells that touch the top and left lines of each square. Do not count those touching the bottom and right lines. Why: the rule assigns every boundary cell to exactly one square and prevents double counting.
  1. Count buds correctly. For budding yeast, count the mother and the bud as two cells if the bud is at least half the size of the mother. Count them as one if the bud is smaller. Why: small buds have not yet committed to division and counting them as separate cells overestimates the population.
  1. Count the duplicate chamber. Use the same squares and the same boundary rule. Why: agreement within 10 to 15% between chambers is the main internal quality check.
  1. Record and calculate. Average the two chamber counts and apply the formula in the next section. Why: recording raw counts before calculating preserves the audit trail if a number looks wrong later.
  1. Clean the slide immediately. Rinse with distilled water, wipe with 70% ethanol, and dry with lens paper. Why: dried sample etches the ruling and attracts debris that mimics cells on the next use.

The Calculation, Worked

The general formula is:

Cells per mL = (total cells counted ÷ number of 1 mm² squares counted) × dilution factor × 10,000

The 10,000 is the reciprocal of the volume of one 1 mm² square at 0.1 mm depth in milliliters (0.0001 mL).

If you count the five medium squares inside the central area, the formula becomes:

Cells per mL = (cells counted ÷ 5) × 25 × dilution factor × 10,000 = cells counted × 50,000 × dilution factor

Worked example

A yeast culture is diluted 1:10 in PBS and mixed 1:1 with 0.4% trypan blue, giving a total dilution factor of 20. You count the full central 1 mm² area in both chambers. Chamber 1 gives 148 cells, chamber 2 gives 132 cells. The mean is 140 cells.

Cells per mL = 140 × 20 × 10,000 = 2.8 × 10⁷ cells per mL.

Of the 140 cells, 120 are unstained and 20 are blue. Viability is 120 ÷ 140 = 85.7%. Viable cells per mL = 2.8 × 10⁷ × 0.857 = 2.4 × 10⁷.

In log form, the total count is log₁₀(2.8 × 10⁷) = 7.45. Converting raw counts to log₁₀ is standard practice in enumeration work because it normalizes the variance across the growth curve and makes successive timepoints comparable [4].

From counts to growth rate

Take two timepoints in exponential phase, t₁ and t₂, with cell densities N₁ and N₂. The specific growth rate µ is:

µ = (ln N₂ - ln N₁) ÷ (t₂ - t₁)

The generation time, or doubling time, is:

g = ln 2 ÷ µ = 0.693 ÷ µ

Generation time is a property of the population, not of a single cell, and it varies with the age structure of the culture [5]. The relation between lag time and generation time is also not fixed. In food microbiology, lag time and generation time respond differently to temperature and other stresses, which is why both parameters are estimated separately rather than derived from one another [6]. In supragingival dental plaque, generation times measured during the early phases of biofilm formation differ from the doubling times seen in liquid batch culture [7]. A hemocytometer growth curve captures the same population-level behavior, provided you sample often enough during exponential phase.

A note on sampling frequency

For a doubling time of 90 minutes, sample every 30 to 45 minutes during exponential phase. For a doubling time of 20 minutes, sample every 10 to 15 minutes or you will miss the transition into stationary phase. Plot log₁₀ cells per mL against time and fit the exponential segment by linear regression. Points outside the exponential window bias the slope and the generation time.

Controls and Quality Assurance

Four controls separate a reliable count from a plausible number.

Duplicate chambers. The two sides of the slide are independent aliquots. A difference greater than 10 to 15% signals a mixing, pipetting, or charging problem. Remix and recount before trusting the average.

Blank chamber. Fill one chamber with sterile PBS or diluent only. The count should be zero or near zero. Debris in the blank points to a dirty slide, contaminated diluent, or precipitate in the dye.

Calibration beads. Count a suspension of certified latex beads of known concentration, typically 10 µm diameter. Recovering the stated concentration within 10% confirms that the chamber depth, the ruling, and your technique are all correct. Run this check when a new slide is opened and whenever a count looks implausible.

Independent method. Pair the hemocytometer count with an OD reading or a plate count at one or two timepoints. Optical density at 600 nm is fast and inexpensive, and modern microplate readers can hold the relative standard deviation of OD measurements below a few percent over a 12-hour run [1]. Agreement between the two methods across the growth curve is stronger evidence than either method alone. Impedimetric sensors offer a third independent readout by tracking the change in electrical resistance that accompanies bacterial metabolism in real time [2].

Expected Results from Hemocytometer Microbial Growth Counts

A correctly executed growth curve has three recognizable phases.

Lag phase. Cell numbers stay flat or rise very slowly while cells adapt to the medium. Hemocytometer counts during lag can be noisy because the cells are enlarging without dividing, and a count based on particle number will rise slightly before the first division.

Exponential phase. Log₁₀ cell number rises linearly with time. The slope gives µ, and the generation time follows from 0.693 ÷ µ. For Escherichia coli in rich medium at 37 °C, the textbook doubling time is about 20 minutes. For Saccharomyces cerevisiae in rich medium at 30 °C, it is about 90 minutes to 2 hours. Slow-growing organisms such as mycobacteria double in 12 to 24 hours.

Stationary phase. The count plateaus. This plateau is where the hemocytometer and the plate count diverge most sharply. The hemocytometer keeps counting dead cells and intact cell walls, so the total particle count stays high. A plate count falls as viability drops. Without a viability dye, a stationary-phase hemocytometer curve overstates the living population.

Compare the hemocytometer total count with a CFU count on the same sample at one or two timepoints. The ratio between them is the CFU-to-particle ratio, and it is a useful correction factor for that organism and medium. Enumeration studies in food and environmental microbiology routinely report viable counts in CFU per gram or CFU per mL, converted to log₁₀ for analysis, and the same convention applies when you report hemocytometer counts alongside them [8][9][10][11][12][13][14]. In liquid culture work, heterotrophic bacterial densities are often maintained at a target such as 1 × 10⁷ CFU per mL, and a hemocytometer can confirm the total particle density between plate counts [14].

Troubleshooting

SymptomLikely causeFix
Count is much lower than expectedCells settled before sampling, or dilution errorVortex immediately before drawing, recheck the dilution arithmetic
Count is much higher than expectedDebris counted as cells, dye precipitate, or multiple focal planes countedUse phase contrast, filter the dye, focus once and count only that plane
Chambers disagree by more than 15%Poor mixing, uneven filling, or a loose coverslipRemix, reseat the coverslip until Newton's rings appear, recharge both sides
Bubbles in the chamberPipetting too fast, or coverslip lifted during fillingCharge slowly at the notch, do not press the coverslip after filling
Cells all in one area of the chamberSlide not level, or insufficient settling timeLevel the bench, wait 3 minutes before focusing
Chamber will not fillNotch blocked by dried sample, or too little volumeClean the notch, use 10 to 15 µL
Counts drift upward during countingEvaporation concentrates the sampleCount within 5 minutes of charging, humidify the room
Trypan blue background too darkDye too concentrated, or cells exposed too longUse 0.2% final, count within 5 minutes
Cells clumpedInsufficient dispersion, or divalent cations bridging cellsVortex longer, add EDTA if the organism tolerates it, filter through a 40 µm mesh
Small bacteria cannot be resolvedWrong chamber depth or too little magnificationSwitch to a Petroff-Hausser chamber and a 100x oil objective with phase contrast

Variations on the Basic Method

Petroff-Hausser chamber. A 0.02 mm deep chamber designed for bacteria. The shallower depth improves resolution and reduces the number of focal planes. The multiplier for the full 1 mm² area is 50,000.

Bürker and Fuchs-Rosenthal chambers. Different rulings and depths, used mainly for eukaryotic cells. Always confirm the depth and the ruling before applying a multiplier.

Disposable plastic hemocytometers. Single-use slides with an integrated coverslip and a certified depth. They cost more per count but eliminate the cleaning and breakage problems of glass, and they are useful for high-containment work.

Automated image-based counters. Instruments that photograph the chamber and identify cells by size and shape apply the same counting principle with less operator fatigue. They improve consistency across operators but require validation against manual counts for each new organism.

Flow cytometry. Counts particles by light scatter and fluorescence at high throughput. It is the method of choice when you need to count thousands of samples or when you need to sort live from dead cells with multiple fluorescent markers.

Viability dyes. Trypan blue and methylene blue are the inexpensive standards. Propidium iodide and SYTO 9 with propidium iodide give sharper live-dead discrimination under fluorescence but require a fluorescence microscope.

Plate counting as the reference. When the question is viability rather than particle number, plate counting remains the reference method, and the injured-cell problem is best addressed with methods designed for it [3]. Hemocytometer counts complement plate counts, they do not replace them.

Storage and Stability Notes

Clean the hemocytometer with distilled water and 70% ethanol after every use, dry it with lens paper, and store it in its case. Do not leave ethanol on the ruling, because prolonged contact can dissolve the optical cement on some slides. Store coverslips dry and discard any with scratches, because a scratched coverslip creates uneven capillary flow.

Trypan blue stock is stable at room temperature for months when protected from light. Discard it if a precipitate forms, because dye crystals are the same size as small cells and will be counted as cells. Methylene blue is stable under the same conditions.

Fixed samples in 2 to 5% formalin can be held at 4 °C for several days before counting. Ethanol-fixed samples keep longer. Unfixed samples should be counted within 30 minutes of collection, because cells continue to divide, change size, and settle.

A charged chamber is stable for about 5 minutes. Evaporation begins at the edges first, which pulls cells toward the coverslip margin and creates a false gradient. If the count takes longer than 5 minutes, discard the slide and recharge.

Common Mistakes and Limitations

The common mistakes in hemocytometer microbial growth work cluster into arithmetic, technique, and interpretation.

Arithmetic mistakes. Using 10,000 when you counted only five medium squares (the correct multiplier is 50,000). Forgetting the dye dilution factor of 2 when trypan blue is mixed 1:1. Forgetting the PBS dilution factor entirely. Reporting cells per mL when the question asked for cells per gram of wet weight. Each of these produces a number that looks believable and is wrong by a known factor.

Technique mistakes. Not mixing before sampling. Charging an overfull or underfull chamber. Counting cells that touch all four boundaries of a square. Counting bubbles, dye crystals, or debris as cells. Reusing a scratched coverslip. Counting only one chamber and calling it a duplicate. Counting too few cells, which inflates the Poisson error above 20%.

Interpretation mistakes. Treating the hemocytometer count as a viable count. Ignoring clumping, which makes one particle stand for several cells. Reporting more significant figures than the method supports. A count of 140 cells in a 1 mm² square carries a sampling error of roughly 8 to 10%, so reporting 2.834 × 10⁷ cells per mL overstates precision. Report two significant figures and the appropriate log₁₀ value.

Limitations of the method itself. The hemocytometer cannot distinguish species, cannot assess metabolic state without a dye, and cannot resolve cells below about 1 µm in brightfield. It is subjective, operator-dependent, and slow compared with automated counters. It counts clumps as single particles. It does not distinguish viable but non-culturable cells from dead cells unless a viability dye is used, and even then the dye reports membrane integrity rather than reproductive capacity. For questions about growth kinetics and predictive modeling, plate counts and optical density remain the dominant data sources, and hemocytometer counts are a complement rather than a substitute [1][15][16].

Individual research or clinical decisions about a specific organism, sample type, or patient still need a qualified microbiologist or veterinarian. The method described here is a measurement technique, not a diagnostic conclusion.

Frequently Asked Questions

Can you use a hemocytometer to measure bacterial growth?

Yes, for bacteria large enough to resolve under light microscopy and preferably with phase contrast. Small rods and cocci are better counted in a Petroff-Hausser chamber at 0.02 mm depth or by plate count.

What is the 10,000 multiplier in a hemocytometer calculation?

It is the reciprocal of the volume of one 1 mm² square at 0.1 mm depth, which is 0.0001 mL. If you count a smaller area, the correct multiplier changes.

Do hemocytometer counts include dead cells?

Yes, by default. The chamber counts every particle of the right size. Add trypan blue, methylene blue, or another viability dye to separate live from dead cells.

How many cells should you count for a reliable result?

Aim for 100 to 200 cells total across the counted squares, which is roughly 20 to 50 cells per 1 mm² square. Fewer cells increases sampling error, and more cells increases overlap.

How long does a hemocytometer count take?

About 25 to 40 minutes of hands-on time for a duplicate count, including sample preparation, charging, settling, counting, and calculation.

Should you count both chambers of the hemocytometer?

Yes. The two chambers are independent aliquots, and agreement within 10 to 15% is the main internal quality check.

Can you convert a hemocytometer count directly to CFU?

No. The hemocytometer gives a total particle count, and CFU measures viable colony-forming units. The two differ by the dead-cell fraction, the clumping fraction, and the viable-but-non-culturable fraction.

Can you store a charged hemocytometer slide before counting?

Only briefly. A charged chamber is stable for about 5 minutes before evaporation distorts the distribution of cells. Recount with a fresh sample if counting is delayed.

Sources

  1. Real-time microbial growth curve (RMGC) system: an improved microplate reader with a graphical interface for automatic and high-throughput monitoring of microbial growth curves.
  2. An impedimetric biosensor for real-time monitoring of bacterial growth in a microbial fermentor
  3. A Novel Double Subculture Method and Its Theory for the Enumeration of Injured Cells in Stressed Microbial Population.
  4. Microbial Growth Study on Pork Loins as Influenced by the Application of Different Antimicrobials
  5. A theory for the age and generation time distribution of a microbial population
  6. Relation between the generation time and the lag time of bacterial growth kinetics.
  7. Microbial generation time during the early phases of supragingival dental plaque formation.
  8. Impact of Refrigerated Storage on Microbial Growth, Color Stability, and pH of Turkey Thigh Muscles
  9. Comparative Analysis of Microbial Growth in Raw and Pasteurized Cow and Camel Milk During Chilled Storage
  10. Development and Validation of a Novel Gas-Washing Bottle Incubation System (GBIS) for Monitoring Microbial Growth in Liquid Media Under Well-Controlled Modified Atmosphere Conditions
  11. Effect of Humidity-Controlled Dehydration on Microbial Growth and Quality Characteristics of Fresh Wet Noodles
  12. Antimicrobial edible coating application of Kecombrang flower concentrate to reduce microbial growth on gourami fish sausage
  13. Ozone application to extend shelf life of vegetables by microbial growth inhibition
  14. Evaluation of biofloc generation protocols to adopt high density nursery rearing of Penaeus vannamei for better growth performances, protective responses and immuno modulation in biofloc based technology
  15. Prediction of population behavior of Listeria monocytogenes in food using machine learning and a microbial growth and survival database
  16. Microbial Growth in Planktonic Conditions