HPLC Liquid Chromatography: Principles and Steps
By Dr. Zubair Khalid, DVM, MS, PhD ·

High performance liquid chromatography, usually shortened to HPLC liquid chromatography, is the workhorse separation technique of modern analytical labs. It takes a dissolved mixture, pushes it through a packed column under high pressure, and lets each compound exit at a slightly different time so you can identify and quantify it. If you can run an HPLC correctly, you can measure drug concentrations in plasma, pesticide residues in food, sugars in seaweed, or impurities in a manufactured tablet.
By the end of this guide you will be able to set up an HPLC system, inject a sample, run a separation, read the chromatogram, and judge whether the result is good enough to trust. You will also understand the difference between normal-phase and reversed-phase chromatography, why reversed-phase dominates pharmaceutical and veterinary drug analysis, and how to avoid the two most common beginner failures: a column that was never equilibrated and a mobile phase that was never degassed.
What you need on hand before you start: an HPLC instrument with a solvent delivery pump, an autosampler or manual injector, a thermostatted column compartment, a detector (UV, fluorescence, or mass spectrometer), and a data system. You also need HPLC-grade solvents, a filtered and degassed mobile phase, a suitable analytical column, sample vials, and a validated method or a starting method to optimize.
What HPLC Liquid Chromatography Actually Does
Chromatography separates compounds based on how they partition between two phases. One phase is stationary (a packed bed of particles inside a column). The other is mobile (a liquid that flows through that bed). A compound that prefers the stationary phase moves slowly. A compound that prefers the mobile phase moves quickly. The difference in travel speed is what separates them.
HPLC improves on classic gravity-fed liquid chromatography by forcing the mobile phase through the column at high pressure, often 100 to 400 bar. Higher pressure lets you use smaller stationary-phase particles, which gives sharper peaks and faster runs. A modern reversed-phase method can separate a drug and its impurities in a single injection, which is exactly what regulated pharmaceutical testing demands [1].
The technique has spread far beyond pharmaceuticals. HPLC supports biochemistry, food testing, environmental monitoring, and materials science because it combines quantitative accuracy, speed, selectivity, and sensitivity in one platform [2]. It is used for therapeutic drug monitoring of first-line antituberculosis drugs in human plasma [3], for measuring monosaccharides and uronic acids in brown seaweed [4], and for detecting tryptophan and serotonin in bovine blood [5].
The Core Steps of an HPLC Run
Every HPLC run follows the same sequence. The details change, but the logic does not.
- Mobile phase preparation. Mix HPLC-grade solvents, buffer to the target pH, filter, and degas.
- Column equilibration. Pump mobile phase through the column until the baseline and backpressure stabilize.
- Sample injection. The autosampler pulls a fixed volume from a vial and injects it into the flowing stream.
- Mobile phase delivery. The pump pushes the mobile phase through the column at a set flow rate.
- Separation. Compounds interact with the stationary phase and separate as they travel.
- Detection. The detector records a signal as each compound elutes.
- Data integration. Software converts the detector signal into peak areas and retention times.
- System suitability check. You confirm resolution, peak shape, and repeatability meet acceptance criteria.
Step 1: Mobile Phase Preparation and Degassing
The mobile phase is the liquid that carries your sample. Its composition controls how strongly compounds bind to the column, so it is the single most powerful variable you can tune.
For reversed-phase work, the mobile phase is usually a polar solvent (water or buffer) mixed with a less polar organic solvent (acetonitrile or methanol). A typical isocratic method might use acetonitrile and ammonium buffer [1]. A stability-indicating method for oxcarbazepine used 75% potassium phosphate monobasic buffer at pH 5.0 and 25% acetonitrile at 1.0 mL/min [6].
Dissolved gas is the enemy here. Gas bubbles form in the pump, cause pressure spikes, and create baseline noise. Degassing removes them. You can degas by helium sparging, vacuum filtration, or an inline degasser. Always filter buffers through a 0.22 or 0.45 micrometer membrane before use. Particulates wear pump seals and plug frits.
Buffer pH matters more than beginners expect. A method for tryptophan and serotonin tested mobile phases at pH 3.5, 4.0, and 4.5 and found pH 3.5 with a methanol to buffer ratio of 20:80 gave the best result [5]. Small pH changes shift the ionization state of analytes and change retention.
Step 2: Column Equilibration
Equilibration means pumping mobile phase through the column until the stationary phase, the mobile phase, and the column temperature reach a steady state. If you skip this, retention times drift and your first few injections are unusable.
A practical rule is to pass at least 10 column volumes of mobile phase before the first injection. For a 4.6 by 250 mm column with 5 micrometer particles, the void volume is roughly 2.5 mL, so 10 column volumes is about 25 mL, or 25 minutes at 1.0 mL/min. Watch the baseline and the system pressure. Equilibration is done when both are flat and stable.
Step 3: Sample Injection
The injector introduces a precise volume of sample into the mobile phase stream. Typical injection volumes range from 1 to 100 microliters. A review of HPLC methods for antituberculosis drugs found injection volumes and flow rates varied widely across validated methods, with flow rates from 0.8 to 1.5 mL/min [3].
Two injection modes exist. Full-loop injection fills a fixed loop completely, giving the best volume reproducibility. Partial-loop injection uses a syringe to meter a smaller volume into the loop, which is more flexible but slightly less precise. Autosamplers handle both automatically and are standard in regulated labs.
Sample preparation matters as much as the instrument. Many biological samples need cleanup before injection. A fabric phase sorptive extraction method extracted five antidepressant drugs from human blood serum without a protein precipitation step and detected them by reversed-phase HPLC with diode array detection at 235 nm [7]. Cleaner samples mean fewer interfering peaks and longer column life.
Step 4: Mobile Phase Delivery by Pump
The pump is the heart of the system. It must deliver a constant, pulseless flow at high pressure. Two modes exist.
Isocratic elution holds the mobile phase composition constant for the whole run. It is simple, reproducible, and ideal when your compounds have similar retention. The oxcarbazepine method used isocratic elution [6].
Gradient elution changes the mobile phase composition over time, usually increasing the organic solvent percentage. It separates compounds with a wide range of polarities in one run. The antidepressant extraction method used gradient elution with diode array detection [7].
Gradient methods need a re-equilibration step between injections. After the gradient finishes, you must return to the starting composition and equilibrate again, or retention times will drift.
Step 5: Separation on the Stationary Phase Column
The column contains the stationary phase, a packed bed of porous particles. Most particles are silica-based and chemically modified with a bonded phase. The bonded phase determines what the column retains.
Separation happens because different compounds spend different fractions of their time adsorbed to the stationary phase versus dissolved in the mobile phase. The retention time is the time from injection to the peak maximum. It is the primary qualitative identifier in HPLC.
Retention depends on several factors: the chemistry of the stationary phase, the composition and pH of the mobile phase, the column temperature, and the flow rate. Change any one and retention shifts.
Step 6: Detection
The detector converts the presence of a compound in the eluent into an electrical signal. Three detectors dominate.
UV or diode array detection (DAD) measures absorbance of ultraviolet or visible light. It is robust, inexpensive, and works for any compound with a chromophore. The antidepressant method used DAD at 235 nm [7], and the oxcarbazepine method used UV detection at 257 nm [6].
Fluorescence detection (FLD) measures emitted light after excitation. It is far more sensitive than UV but only works for naturally fluorescent compounds or those that can be derivatized. A method for tryptophan and serotonin in bovine blood used HPLC-FLD and achieved intraday precision of 2.20% for tryptophan and 2.36% for serotonin [5]. A teaching experiment coupled online solid phase extraction with HPLC and post-column derivatization with fluorescence detection to measure trace carbamate pesticides in vegetables [8].
Mass spectrometry (MS) measures the mass-to-charge ratio of ionized molecules. It gives both quantification and structural identification and is the most sensitive option. Reversed-phase HPLC coupled to tandem mass spectrometry has been used for enantioselective determination of cyclaniliprole in plant-derived foods [9] and for detecting pancreatic stimulants in equine urine [10].
Step 7: Data Integration
Integration converts the raw detector trace into numbers. The software identifies peak start and end points, draws a baseline, and calculates peak area, peak height, retention time, and peak width.
Peak area is proportional to the amount of compound, so it is the basis for quantification. You build a calibration curve by injecting standards of known concentration and plotting area against concentration. Then you interpolate unknown sample areas against that curve.
Integration parameters matter. Wrong baseline placement inflates or deflates peak areas. Always inspect the integration visually before accepting results.
Normal-Phase vs Reversed-Phase Chromatography
The terms normal-phase and reversed-phase describe the relative polarity of the stationary and mobile phases. Getting this distinction right is essential because the two modes retain completely different compound classes.
In normal-phase chromatography, the stationary phase is polar (bare silica, alumina, or an amino-bonded phase) and the mobile phase is nonpolar (hexane, heptane, or dichloromethane, often with a small amount of a polar modifier like 2-propanol). Polar analytes bind strongly to the polar stationary phase, so they elute late. Nonpolar analytes elute early. Normal-phase is used for separating isomers, lipids, and compounds that are unstable in water.
In reversed-phase chromatography, the stationary phase is nonpolar (most commonly a C18 or octadecylsilane bonded phase, sometimes C8) and the mobile phase is polar (water or buffer mixed with acetonitrile or methanol). Nonpolar analytes bind strongly to the nonpolar stationary phase, so they elute late. Polar analytes elute early. Reversed-phase dominates pharmaceutical, veterinary, and food analysis because most drug molecules and many natural products are moderately nonpolar and dissolve well in water-organic mixtures.
| Feature | Normal-Phase | Reversed-Phase |
|---|---|---|
| Stationary phase polarity | Polar (silica, amino) | Nonpolar (C18, C8) |
| Mobile phase polarity | Nonpolar (hexane, heptane) | Polar (water, buffer, acetonitrile) |
| Elution order | Nonpolar first, polar last | Polar first, nonpolar last |
| Typical analytes | Lipids, isomers, fat-soluble vitamins | Drugs, metabolites, pesticides, peptides |
| Typical use | Isomer separation, nonaqueous samples | Pharmaceutical, veterinary, food, clinical |
Reversed-phase HPLC is the default for pharmaceutical and veterinary drug analysis. A review of HPLC methods for therapeutic drug monitoring of first-line antituberculosis drugs found that studies used C18 and C8 stationary phases with a range of mobile phases and both gradient and isocratic elution [3]. A stability-indicating method for oxcarbazepine used a Supelco Discovery C18 column (4.6 by 250 mm, 5 micrometer particles) with a phosphate buffer and acetonitrile mobile phase [6]. A method for pentacyclic triterpenes in Centella asiatica used a YMC-Pack ODS-A column with a water and acetonitrile gradient containing formic acid and beta-cyclodextrin as a mobile-phase additive [11].
Reading a Chromatogram: Retention Time, Resolution, and Peak Asymmetry
Three parameters tell you whether a separation is working. Learn to read them before you trust any result.
Retention Time
Retention time (tR) is the time from injection to the peak maximum. It identifies a compound under fixed conditions. If you change the method, retention time changes, so you cannot compare retention times across different methods.
Retention time also depends on column equilibration. A column that is not fully equilibrated gives drifting retention times, which is why equilibration is not optional.
Resolution
Resolution (Rs) measures how well two adjacent peaks are separated. The formula is:
Rs = 2 * (tR2 - tR1) / (w1 + w2)
Where tR1 and tR2 are the retention times of the two peaks, and w1 and w2 are their baseline widths.
A resolution of 1.5 or greater is the standard acceptance criterion for baseline separation in pharmaceutical method validation. Below 1.5, peaks overlap and quantification becomes unreliable. A method for pentacyclic triterpenes achieved resolution values of 2 or greater after optimization with beta-cyclodextrin as a mobile-phase additive [11].
To improve resolution, you can increase column length, reduce particle size, lower the flow rate, adjust the mobile phase composition, or change the column temperature.
Peak Asymmetry
Peak asymmetry (also called tailing factor or asymmetry factor, As) measures how symmetric a peak is. A perfectly symmetric peak has an asymmetry factor of 1.0. Most acceptance criteria allow 0.8 to 1.5, with some methods allowing up to 2.0.
Tailing peaks (As greater than 1.5) usually mean secondary interactions with the stationary phase, often with residual silanol groups on the silica surface. A study on deep eutectic solvents as mobile-phase additives found that they improved peak shapes and resolution in reversed-phase separation of G-quadruplexes, partly by shielding silanol groups on the C18 stationary phase [12].
Fronting peaks (As less than 0.8) usually mean column overload or a solvent mismatch between the sample and the mobile phase.
A Beginner-Ready HPLC Run Protocol
This protocol assumes a reversed-phase isocratic method with UV detection. Adjust volumes and times to your specific column and method.
Before you start:
- Prepare mobile phase: filter through 0.22 micrometer membrane, degas by vacuum or helium sparging.
- Check solvent levels in all reservoirs.
- Check waste container has space.
Step 1: Equilibrate the column
- Set flow rate to 1.0 mL/min.
- Pump mobile phase for at least 10 column volumes (about 25 minutes for a 4.6 by 250 mm column).
- Monitor baseline and pressure until both are stable.
Step 2: Set detector wavelength
- For UV detection, set to the absorbance maximum of your analyte. Example: 257 nm for oxcarbazepine [6], 235 nm for antidepressant drugs [7].
Step 3: Prepare samples and standards
- Filter samples through 0.22 micrometer syringe filter.
- Prepare calibration standards at known concentrations.
- Load vials into autosampler tray.
Step 4: Run system suitability injections
- Inject the system suitability standard at least five times.
- Check retention time repeatability (relative standard deviation should be low, typically under 2%).
- Check resolution between critical peak pairs (must be 1.5 or greater).
- Check peak asymmetry (typically 0.8 to 1.5).
Step 5: Run calibration standards
- Inject each calibration standard.
- Build calibration curve by plotting peak area against concentration.
Step 6: Run samples
- Inject each sample.
- Record retention time and peak area.
Step 7: Integrate and review
- Inspect each chromatogram visually.
- Verify baseline placement.
- Confirm peaks are within the calibration range.
Step 8: Shut down
- Flush column with appropriate storage solvent (usually high organic content for reversed-phase).
- Reduce flow rate gradually before stopping the pump.
- Cap column ends if removing from the instrument.
Worked Example: Calculating Resolution
Suppose you have two peaks. Peak 1 has a retention time of 5.20 minutes and a baseline width of 0.30 minutes. Peak 2 has a retention time of 5.85 minutes and a baseline width of 0.32 minutes.
Rs = 2 * (5.85 - 5.20) / (0.30 + 0.32)
Rs = 2 * 0.65 / 0.62
Rs = 1.30 / 0.62
Rs = 2.10
A resolution of 2.10 exceeds the 1.5 acceptance criterion, so this separation is acceptable.
Now suppose the same peaks had widths of 0.45 minutes each.
Rs = 2 * 0.65 / (0.45 + 0.45)
Rs = 1.30 / 0.90
Rs = 1.44
A resolution of 1.44 falls below 1.5, so you would need to improve the separation before trusting quantitative results.
Common Mistakes and Limitations
Skipping column equilibration. This is the most common beginner error. If you inject before the baseline and pressure stabilize, retention times drift and your first injections are wasted. Always allow at least 10 column volumes of equilibration.
Forgetting to degas the mobile phase. Dissolved gas causes pump cavitation, pressure spikes, and baseline noise. Degas every mobile phase, every time. Buffers are especially prone to gas buildup.
Using the wrong mobile phase pH. Small pH changes shift retention dramatically for ionizable compounds. Always measure pH after mixing organic and aqueous components, because pH shifts when solvents combine.
Overloading the column. Injecting too much sample causes fronting peaks and poor resolution. Reduce injection volume or dilute the sample.
Ignoring peak asymmetry. Tailing peaks usually mean secondary interactions or a deteriorating column. Check the asymmetry factor on every run.
Mismatching sample solvent and mobile phase. If your sample is dissolved in a much stronger solvent than the mobile phase, peaks distort. Dissolve samples in mobile phase or a weaker solvent when possible.
Not filtering samples. Particulates plug frits and damage the column. Filter every sample through a 0.22 micrometer membrane.
Assuming retention time alone identifies a compound. Retention time is not definitive identification. Confirm with a second method, a standard addition, or a detector that gives structural information.
Limitations of the technique itself. HPLC cannot separate compounds that have identical interactions with the stationary phase. It requires method development for each new analyte class. It consumes solvents, though green chemistry approaches are reducing that footprint [13]. And it needs regular maintenance: pump seals wear, columns degrade, and detectors drift.
Individual samples and unusual matrices may need expert review. A veterinarian or analytical chemist should be consulted for clinical or regulatory decisions.
How the Steps Fit Together
The workflow below shows the main decision path from sample to validated result.
flowchart TD
A[Prepare mobile phase] --> B[Degas and filter]
B --> C[Equilibrate column]
C --> D[Inject sample]
D --> E[Pump delivers mobile phase]
E --> F[Separation on column]
F --> G[Detector records signal]
G --> H[Integrate peaks]
H --> I{Resolution 1.5 or greater}
I -->|Yes| J[Report result]
I -->|No| K[Adjust method]
K --> C
Method Development and Optimization
Beginners often inherit a validated method and run it as written. That is the right approach for routine work. But understanding how methods are developed helps you troubleshoot when something goes wrong.
Method development starts with screening. You test a few column chemistries, mobile phase compositions, and pH values to find conditions that separate your analytes. A stability-indicating method for itraconazole used a Design of Experiments approach to optimize chromatographic parameters and quantify the drug and its impurities in a single injection [1]. A method for genotoxic impurities in rabeprazole used Quality by Design principles, with a switching valve to divert the main drug peak to waste while the impurities were detected [14].
Optimization targets resolution, run time, and robustness. You want baseline separation (resolution 1.5 or greater) in the shortest practical run time, with a method that tolerates small variations in temperature, pH, and flow rate.
Modern method development also considers environmental impact. Green analytical chemistry encourages replacing toxic solvents, reducing solvent consumption, and using miniaturized or energy-efficient instrumentation [13]. High-temperature liquid chromatography is one green approach that uses pure water as the mobile phase at elevated temperatures, though it requires stationary phases stable at 60 to 90 degrees Celsius [15].
Detector Selection in Practice
Choosing a detector depends on what you need to measure and how sensitive you need to be.
UV and DAD are the default for pharmaceutical analysis. They are robust, well understood, and adequate for most drug assays. The antidepressant method used DAD at 235 nm and achieved a limit of detection of 0.15 nanograms per microliter [7].
Fluorescence is the choice when you need maximum sensitivity for fluorescent compounds. The tryptophan and serotonin method achieved mean accuracy of 105.2% for tryptophan and 99.4% for serotonin with precision under 5% [5].
Mass spectrometry is the choice when you need identification, confirmation, or trace-level quantification in complex matrices. Reversed-phase HPLC with tandem mass spectrometry has been used for enantioselective determination of cyclaniliprole in fruits, vegetables, and cereals [9] and for detecting acetohexamide and chlorpropamide in equine urine [10].
Evaporative light scattering detection (ELSD) is useful for compounds without a chromophore, such as sugars. A method for monosaccharides, mannitol, and uronic acids in brown seaweed used HPLC-ELSD with hydrophilic interaction liquid chromatography conditions and achieved average recovery of 102% across all compounds [4].
Column Chemistry and Stationary Phase Selection
The stationary phase determines what your column can separate. Most HPLC columns use silica particles with a bonded phase.
C18 (octadecylsilane) is the most common reversed-phase stationary phase. It retains moderately nonpolar compounds well and works for a wide range of drugs and natural products. The oxcarbazepine method used a C18 column [6], and the antidepressant method used reversed-phase separation [7].
C8 (octylsilane) is slightly less hydrophobic than C18. It is useful when compounds are too strongly retained on C18 or when you need different selectivity. A method for genotoxic impurities in rabeprazole used a C8 column [14].
Mixed-mode stationary phases combine reversed-phase and ion-exchange or other interactions to handle compounds that single-mode columns cannot separate well. They are increasingly used for complex samples with a wide polarity range [2].
Chiral stationary phases separate enantiomers. They are used when a drug has a chiral center and the two enantiomers have different biological activity. Reversed-phase chiral separation has been demonstrated for cyclaniliprole enantiomers using a Superchiral R-IC column [9], and composite chiral stationary phases based on amylose derivatives have been evaluated for enantioseparation [16].
Specialty phases include fluorinated stationary phases for fluorinated analytes [17] and superhydrophobic fluorinated microporous microspheres for separating polar pyrimidines under aqueous mobile phases [18].
System Suitability: The Gatekeeper of Valid Results
System suitability is a set of tests you run before analyzing samples. It confirms the instrument, column, and method are working together correctly. If system suitability fails, sample results are invalid.
Typical system suitability parameters include:
- Resolution between critical peak pairs (acceptance criterion usually 1.5 or greater).
- Retention time repeatability across replicate injections (relative standard deviation typically under 2%).
- Peak area repeatability across replicate injections (relative standard deviation typically under 2%).
- Peak asymmetry (typically 0.8 to 1.5).
- Theoretical plates (a measure of column efficiency, higher is better).
Run system suitability at the start of every batch, after any change to the method, and after column maintenance.
Frequently Asked Questions
What is the difference between HPLC and liquid chromatography?
HPLC is a type of liquid chromatography that uses high pressure to push the mobile phase through a packed column. Classic liquid chromatography relies on gravity. The high pressure in HPLC allows smaller particles and faster, sharper separations.
What does reversed-phase mean in HPLC?
Reversed-phase means the stationary phase is nonpolar (usually C18) and the mobile phase is polar (water or buffer with acetonitrile or methanol). It is called reversed because it flips the polarity arrangement of normal-phase chromatography.
Why is reversed-phase HPLC used for drug analysis?
Most drug molecules are moderately nonpolar and dissolve well in water-organic mixtures. Reversed-phase columns retain them reliably, and the mobile phase is compatible with biological samples. A review of antituberculosis drug monitoring found that validated methods used C18 and C8 columns with a range of polar mobile phases [3].
What is a good resolution in HPLC?
A resolution of 1.5 or greater is the standard acceptance criterion for baseline separation. Below 1.5, peaks overlap and quantification becomes unreliable.
Why does my retention time keep drifting?
The most common cause is insufficient column equilibration. Pump at least 10 column volumes of mobile phase before injecting. Other causes include temperature fluctuation, mobile phase evaporation, and column degradation.
Do I need to degas my mobile phase?
Yes. Dissolved gas causes pump cavitation, pressure spikes, and baseline noise. Degas every mobile phase by vacuum filtration, helium sparging, or an inline degasser.
What is peak asymmetry and why does it matter?
Peak asymmetry measures how symmetric a peak is. Tailing or fronting peaks reduce resolution and quantification accuracy. Most methods accept asymmetry factors between 0.8 and 1.5.
Can HPLC identify unknown compounds?
HPLC with UV detection cannot definitively identify unknowns because retention time alone is not unique. Mass spectrometry or a second orthogonal method is needed for identification.
Related Articles
- Affinity Column Chromatography: Principles and Applications
- Gel Permeation Chromatography: Principles and Applications
- How to Get Purified Gel for Chromatography: A Step-by-Step Guide
- Gel Filtration Chromatography Matrix: Principles and Applications
- Gel Filtration Chromatography (GFC): Principles and Applications
- Gel Filtration Chromatography: Principles and Applications
- Gas Liquid Chromatography: Uses and Principle
- Acid-Fast Stain: Principle, Procedure, and Interpretation
Sources
- Development and Optimisation of a Stability-Indicating Analytical Method for Itraconazole in Drug Product by High-Performance Liquid Chromatography Using a QbD Approach.
- [[Research progress on high performance liquid chromatography stationary phases based on mixed mode].](https://pubmed.ncbi.nlm.nih.gov/42246096/)
- High-Performance Liquid Chromatography for the Therapeutic Drug Monitoring of First-Line Antituberculosis Drugs: A Comprehensive Review.
- High-performance liquid chromatography with evaporative light scattering detection for the determination of monosaccharides, mannitol and uronic acids in brown seaweed.
- Methods for storage and determination of tryptophan and serotonin in bovine blood by high-performance liquid chromatography with fluorescence detection.
- Validated stability indicating reverse phase liquid chromatography method for oxcarbazepine and its degradation products with greenness evaluation.
- An improved fabric phase sorptive extraction method for the determination of five selected antidepressant drug residues in human blood serum prior to high performance liquid chromatography with diode array detection.
- [[Self-designed experiment in instrumental analysis: online solid phase extraction-high performance liquid chromatography- post column derivatization for the determination of trace pesticide residues in vegetables].](https://pubmed.ncbi.nlm.nih.gov/42754943/)
- Novel application of reversed-phase high performance liquid chromatography-tandem mass spectrometry for enantioselective determination of cyclaniliprole enantiomers in plant-derived foods.
- [[Determination of two pancreatic stimulants in equine urine using ionic liquid-based deep eutectic solvent-assisted liquid-liquid microextraction and ultra-high performance liquid chromatography-tandem mass spectrometry].](https://pubmed.ncbi.nlm.nih.gov/42754940/)
- Enhanced selectivity in RP-HPLC-UV separation of structurally similar pentacyclic triterpenes in Centella asiatica using β-cyclodextrin as a mobile-phase additive.
- Effect and Mechanism of Deep Eutectic Solvents as Mobile-Phase Additives on Retention Behavior of G-Quadruplexes in Reversed-Phase High-Performance Liquid Chromatography.
- Green Approaches in High-Performance Liquid Chromatography for Sustainable Food Analysis: Advances, Challenges, and Regulatory Perspectives.
- Analysis of potential genotoxic impurities in rabeprazole active pharmaceutical ingredient via Liquid Chromatography-tandem Mass Spectrometry, following quality-by-design principles for method development.
- Sustainable separation using high-temperature liquid chromatography (HTLC): mechanism, packing materials, challenges, and applications.
- HPLC enantioseparation on composite chiral stationary phases based on homosubstituted and heterosubstituted amylose derivatives.
- Fluorous Affinity-Driven Retention of Fluorinated Analytes on a High-Density Fluorinated Silica Gel Stationary Phase.
- Superhydrophobic fluorinated microporous microsphere stationary phase for liquid chromatography separation of highly polar pyrimidine compounds under water mobile phase.