Flow Cytometry Explained (Video)

Written guide: Flow Cytometry Procedure

This video explains how a flow cytometer measures single cells with lasers, forward and side scatter and fluorescent antibodies, and how gating, compensation and cell sorting work.

Watch on YouTube

Chapters

  1. 0:00 What flow cytometry measures
  2. 0:52 How flow cytometry began
  3. 1:43 Fluidics and hydrodynamic focusing
  4. 2:34 Forward and side scatter
  5. 3:35 Fluorescent labels and detectors
  6. 4:47 Reading the data with gating
  7. 5:57 Compensation and controls
  8. 6:45 Cell sorting
  9. 7:21 Uses in research and medicine
  10. 8:15 Review and question
Video transcript

0:00 What flow cytometry measures

In this video, we will learn about flow cytometry, a method that measures single cells one at a time as they stream past a laser beam. Flow cytometry measures the light that a cell scatters and the fluorescence it gives off, so a single tube of blood yields measurements on many thousands of individual cells within a minute. These measurements tell us how large a cell is, how complex its interior is, and which proteins it carries, so we can count and compare the cell types in a mixed sample. First, we will see how the method began, and then we will follow a cell through the instrument, from the fluid system and light scatter to the fluorescent labels and the detectors. After that, we will learn to read the data with gating and compensation, and we will finish with cell sorting and the main uses of flow cytometry in research and medicine.

0:52 How flow cytometry began

Flow cytometry grew out of a practical need, which was to count blood cells quickly. In 1953, Wallace Coulter patented a counter that drew cells one by one through a small opening in a fluid and counted each cell by the change in electrical resistance it caused. In 1965, Mack Fulwyler at Los Alamos described a machine that measured cells in a stream and separated them into charged droplets. In 1969, Leonard Herzenberg's group at Stanford added fluorescence detection and built the fluorescence-activated cell sorter, or FACS, which Becton Dickinson sold as a commercial instrument from 1974. Scientists agreed on the name flow cytometry at a conference in 1976, from cyto, meaning cell, and metry, meaning measurement.

1:43 Fluidics and hydrodynamic focusing

Before the instrument can measure single cells, its fluid system, or fluidics, has to align the cells in single file. The sample has to be a suspension of single cells, such as blood or dissociated, filtered tissue. The instrument injects the sample into the center of a faster stream of buffer, the sheath fluid, inside a narrow flow cell. The two fluids move in laminar flow, in smooth parallel layers that do not mix. The faster sheath squeezes the sample stream into a thin core only slightly wider than a cell, and we call this hydrodynamic focusing. The cells therefore pass in single file through the spot where the laser crosses the stream, the interrogation point. A higher sample pressure speeds up the run but widens the core and lowers precision, so low pressure suits precise measurements such as DNA content.

2:34 Forward and side scatter

When a cell reaches the laser, the instrument first records how the cell scatters light. Light scattered at small angles along the direction of the beam is forward scatter, or FSC, and a larger cell generally gives a stronger forward scatter signal. Light scattered at about 90 degrees to the beam is side scatter, or SSC, and it increases with internal complexity, such as granules and an irregular nucleus. Neither signal needs a stain, and after red cell lysis, a plot of forward scatter against side scatter already separates three populations of white blood cells. Lymphocytes are small cells with low side scatter, while monocytes are larger and appear further to the right. Granulocytes, such as neutrophils, give the highest side scatter because their cytoplasm is full of granules. In a published plot of isolated blood mononuclear cells, the lymphocytes form a dense cluster low and to the left, and the monocytes sit higher up and further right.

3:35 Fluorescent labels and detectors

Scatter alone cannot tell a helper T cell from a B cell of the same size, so the method adds fluorescent labels that report which proteins a cell carries. Antibodies bind specific surface proteins, which immunologists name with CD numbers, such as CD4 on helper T cells. Each antibody carries a fluorochrome, a dye that absorbs light at one wavelength and emits light at a longer wavelength. The blue 488 nanometer laser excites FITC, which emits green light near 520 nanometers, and the same laser excites PE, which emits yellow-orange light near 575 nanometers. Inside the instrument, dichroic mirrors reflect one range of wavelengths and pass another, and band-pass filters let only a narrow band through, so each color reaches its own detector. The detectors are photomultiplier tubes, which turn weak light signals into an electrical pulse that rises as the cell enters the beam and falls as the cell leaves. The electronics record the height, area and width of the pulse and convert them to numbers. A cell therefore produces one event, a record that holds one value per detector.

4:47 Reading the data with gating

A run usually produces tens of thousands of events, and gating selects the cells of interest from those events. A histogram shows one measurement against the number of events, while a dot plot shows two measurements with one dot per cell, and fluorescence axes usually use a logarithmic scale. A gate is a region drawn around a population, and the next plot shows only the cells inside the region. In a published gating sequence for blood T cells, the first gate selects lymphocytes on a plot of forward against side scatter. The second gate removes doublets, because two cells stuck together give a pulse with a larger area for the same height. A plot of forward scatter area against forward scatter height therefore separates doublets from single cells. A viability dye enters only dead cells, so the third gate keeps the live cells, and the fourth gate selects CD3 positive cells, which are the T cells. Finally, a plot of CD4 against CD8 divides the T cells into quadrants, with CD4 positive helper T cells, CD8 positive cytotoxic T cells, and small double-negative and double-positive groups.

5:57 Compensation and controls

A gate is only trustworthy when the signal in each detector comes from one dye, and compensation corrects the signals toward that goal. Emission spectra are broad, so some FITC emission reaches the PE detector, and we call this effect spillover. Compensation subtracts the expected spillover from each detector, using samples stained with one dye at a time, the single-stain controls. An unstained sample shows the cells' own background fluorescence, which is their autofluorescence. A fluorescence-minus-one control, or FMO, omits one dye, so it shows where to set the gate boundary for that dye. Spectral cytometers instead record the full emission spectrum of each cell and separate the dyes mathematically, and we call this step unmixing.

6:45 Cell sorting

In an analyzer, the measured cells go to waste, while a sorter also collects the selected cells. In a droplet sorter, the stream leaves a vibrating nozzle, so the stream breaks into droplets at a steady rate. When the instrument detects a wanted cell, it charges the stream at the moment the droplet carrying that cell breaks off. The charged droplet passes between two charged deflection plates, which pull the droplet into a collection tube, while uncharged droplets fall to waste. To check the purity of the sort, the operator runs a small sample of the sorted cells through the instrument again.

7:21 Uses in research and medicine

With measurement, gating and sorting in place, we can look at how laboratories use flow cytometry. In HIV infection, the CD4 T cell count per microliter of blood guides treatment, and a count below 200 cells per microliter defines AIDS. Clinicians classify leukemias and lymphomas by the set of CD markers on the abnormal cells, a process we call immunophenotyping. To study the cell cycle, we stain cells with a dye such as propidium iodide, which binds DNA in proportion to its amount. In a published histogram, a large peak of cells sits in G1, a second peak at twice the DNA content marks G2 and mitosis, and cells in S phase lie between the two peaks. Annexin V binds a membrane lipid that appears on the outer surface of cells early in apoptosis, so we combine it with a dead-cell dye to separate live, early apoptotic and dead cells.

8:15 Review and question

Here is a short review of the main points. Hydrodynamic focusing lines cells up in single file, and forward and side scatter report the size and internal complexity of a cell. Fluorescent antibodies, filters and detectors report which proteins a cell carries, while gating and compensation turn the events into cell populations. Sorters then collect the chosen cells in charged droplets. Now try a question on what you learned. Two cells stuck together pass the laser as one event, so which two measurements would you plot to remove these events, and why does the pair work? Please write your answer in the comments below.

Read more