# Biosafety Level 1 and 2 Labs: Common Mistakes in Safety Practices

The most common [biosafety level](/knowledge/diagnostics/emerging-tech/biosafety-levels-1-4-a-comparative-framework-for-diagnostic-laboratory-design) 1 and 2 mistakes are not missing cabinets or broken autoclaves. They are gaps between the written risk assessment and what people actually do at the bench. A laboratory can own every piece of required equipment, post every sign, and still release aerosols during a routine vortex step because the procedure was never assessed for aerosol generation in the first place.

The first thing to check is simple. Pick one procedure performed in the lab today and answer three questions in writing: which agent is involved, at what concentration and volume, and which step creates droplets, aerosols, or sharp injury risk. If the answer is "we follow BSL-2 for everything," the assessment has drifted into a label. A label cannot tell you whether the work belongs in a cabinet, whether a splash shield is enough, or whether the waste stream needs autoclaving before it leaves the room.

This guide covers the failure modes seen most often in biosafety level 1 and 2 labs, the mechanisms behind them, and the fixes that hold up under audit. It is written for laboratory professionals, supervisors, and diagnostic staff who run the bench rather than the paperwork.

## Definitions Used Throughout

A **biosafety level** is a combination of practices, safety equipment, and facility design matched to the risk posed by the agents handled and the procedures performed. A **biological safety cabinet (BSC)** is a ventilated enclosure that protects the worker, the product, or both by moving air through filters and a controlled inflow. An **aerosol** is a suspension of particles small enough to remain airborne and be inhaled, typically under about 5 micrometers. **Personal protective equipment (PPE)** means lab coat, gloves, eye and face protection, and respiratory protection when indicated. A **biological indicator** is a prepared spore strip or ampoule used to confirm that a sterilization cycle actually killed a resistant organism. A **biosafety risk assessment** is the documented process of identifying hazards, exposure routes, and controls for a defined task.

## What BSL-1 and BSL-2 Actually Require

Biosafety level 1 applies to agents not known to consistently cause disease in healthy adults. Work can be done on an open bench with a sink for handwashing, doors that close, routine decontamination, and minimal PPE. Biosafety level 2 applies to agents of moderate hazard that are associated with human disease, with the primary routes of exposure being percutaneous injury, ingestion, and mucous membrane contact. BSL-2 keeps all BSL-1 practices and adds restricted access during work, biohazard signage, a written [biosafety manual](/knowledge/diagnostics/emerging-tech/biosafety-manual-development-a-template-for-diagnostic-laboratories), an autoclave available within the facility or nearby, PPE including a lab coat and gloves, and a BSC for procedures with aerosol potential.

| Element | BSL-1 | BSL-2 |
|--|--|--|
| Bench work | Open bench acceptable | Open bench only for non-aerosol work |
| Cabinet | Not required | Required for aerosol-generating steps |
| Access | Limited as needed | Restricted when work is in progress |
| Signage | Optional | Required at entry with agent and contact |
| Manual | Standard practice | Required, with spill and waste procedures |
| Autoclave | On site preferred | Available in facility or nearby |
| PPE | Minimal | Lab coat, gloves, eye protection |
| Sharps | Standard precautions | Written sharps precautions |

The comparison table hides the real issue. Almost every BSL-2 requirement is conditional. The cabinet is required for *aerosol-generating* steps, not for all steps. Eye protection is required for *splash risk*, not for all tasks. Access restriction applies *when work is in progress*, not permanently. Every conditional creates a judgment call, and every judgment call creates a place where practice can quietly diverge from the manual.

## Why the Same Mistakes Recur

Biosafety level 1 and 2 mistakes persist because the failures are invisible until an exposure occurs. A misdirected pipette that creates a fine mist leaves no spill on the bench. A centrifuge tube that cracks inside a sealed cup produces nothing to clean up if the seal holds, and a full aerosol release if it does not. The feedback loop is broken: bad practice usually produces no immediate consequence, so it gets repeated and normalized.

A second reason is design drift. When [biosafety levels](/knowledge/diagnostics/emerging-tech/biosafety-levels-1-4-a-decision-framework-for-diagnostic-laboratories) are assigned to a room rather than to a task, agents migrate. A procedure originally done in a cabinet gets moved to an open bench because the cabinet is occupied. A new assay arrives with a new agent and the assessment is copied from the previous one. The result is a laboratory that believes it is operating at BSL-2 while a subset of its work is effectively at BSL-1 practice with BSL-2 signage.

A third reason is that facility decisions made early are hard to reverse. Institutional experience with veterinary and diagnostic teaching facilities shows that a hazard- and risk-based strategy should be formulated before architects begin design, and that failing to do so produces costly construction mistakes. The same body of experience found that BSL-2 facilities were best identified from risk scenarios and transmission routes rather than from a generic room list, and that the resulting design successfully prevented spread of infection from isolation areas over five years of operation, with successful elimination of *Salmonella* and MRSA from a large animal clinic [1]. The lesson transfers directly to human clinical and research laboratories: the risk assessment drives the building, and the building then constrains the assessment.

## The Root Causes Behind Biosafety Level 1 and 2 Mistakes

### Knowledge and training gaps

Staff often learn biosafety by copying the person beside them. If the trainer was taught to flame a loop on an open bench, the trainee inherits that behavior. Microbiology biosafety training that consists of an annual video module and a signature does not correct a technique error, because it never observes the technique. Competence is a motor skill and a decision skill, and both require direct observation.

### Behavior and normalization

Once a shortcut survives a month without a consequence, it becomes the local standard. Gloves worn into the hallway, a second pair of gloves used for a phone call, a spill blotted with paper towels and no contact time. The mechanism is social: the shortcut is what competent-looking people do, so it reads as correct.

### Engineering and design failures

A cabinet in a high-traffic corridor, a sink too far from the exit door, a handwash station with no soap, a sharps container mounted above knee height, an autoclave across the building that nobody uses at the end of a shift. Design determines behavior more reliably than policy does.

### Supply and logistics failures

Empty spill kits, expired disinfectant, wrong bleach concentration, gloves in one size only, no eye protection within reach. Staff improvise, and improvisation is where errors enter.

### Documentation and oversight failures

A biosafety manual last revised before the current assays existed. A risk assessment with no date and no author. An autoclave log with identical handwriting across three shifts. Documentation failures make every other failure harder to find.

## Troubleshooting Table: Common Biosafety Level 1 and 2 Mistakes

| Symptom | Likely cause | How to confirm | Fix |
|--|--|--|--|
| No documented risk assessment for a current assay | Assessment copied from an older procedure | Ask staff for the assessment for that specific assay and check the date and agent name | Rewrite by task, include aerosol and sharps steps, review with supervisor |
| Aerosol-generating work done on open bench | Cabinet seen as optional or unavailable | Observe a full run and note vortexing, pipetting, opening of tubes, loop work | Move the step into the cabinet or use a validated closed system, then update the manual |
| Cabinet airflow alarm or uneven sash | Blocked front or rear grille, overloading | Inspect grille for notes, tips, racks, and waste bags | Clear the grille, keep items at least 4 inches behind the front edge, recertify |
| Repeated skin or eye irritation complaints | Disinfectant or PPE mismatch | Review disinfectant concentration and glove type against the safety data sheet | Switch to compatible gloves, reduce concentration, add eye protection |
| Waste bags leaking in the autoclave | Overfilling, wrong bag type, no secondary container | Open the autoclave after a cycle and inspect the load | Fill bags to two-thirds, use polypropylene secondary pans, validate cycle |
| Autoclave indicator tape changes color but staff unsure of sterility | Tape is a process indicator, not a sterilization proof | Compare tape only against biological indicator results | Run biological indicators on a defined schedule and log results |
| Sharps container overfilled | No replacement schedule | Check fill lines at each shift change | Replace at three-quarters full, place containers at reachable height |
| Spill kit missing or expired | No assigned owner | Open the kit and check contents against the inventory list | Assign monthly check, post inventory card, restock immediately |
| Gloves worn into corridors and on phones | Habit and unclear boundaries | Observe a shift transition and door use | Set a glove-off threshold at the lab door, provide phone pouches |
| Training records exist but new staff make technique errors | Training not competency-based | Watch a new hire perform a mock spill and a cabinet run | Add observed competency, sign-off, and annual practical check |
| Centrifuge opened immediately after stop | Speed of workflow | Watch a run from stop to lid opening | Wait at least 10 minutes at BSL-2 and 30 minutes at BSL-3 before opening |
| Equipment sent for repair without decontamination | No decontamination certificate | Ask the service provider for the certificate on file | Add a decontamination form to the work order process |

## The Most Common Mistakes in Detail

### 1. A risk assessment that no longer matches the work

This is the upstream error that produces most of the downstream ones. A usable biosafety risk assessment is procedure-specific, not agent-specific alone. It states the agent, the concentration, the volume, the route of exposure, the aerosol potential of each step, the sharps involved, the PPE required, the engineering control used, the decontamination method, and the waste path. It carries a date and an author and is reviewed when a procedure, reagent, or piece of equipment changes.

The failure mechanism is copy-forward. Someone drafts an assessment for an organism, then a new assay with higher volume or a more vigorous mixing step is added without a new review. The document stays technically true for the old workflow and silently false for the new one. Fix this by tying each assessment to a named procedure identifier and requiring a new review whenever the procedure changes.

### 2. Working outside the biological safety cabinet when the step generates aerosols

The BSC is required for procedures with aerosol potential, not for every manipulation. The mistake runs in both directions: some labs put everything in the cabinet out of caution and then overload it, and others keep high-aerosol steps on the bench. High-risk steps include vortexing with the cap loose, vigorous pipetting, opening tubes after centrifugation, flame-loop work, and any manipulation that produces a visible mist or a fine spray.

Mechanism: droplets small enough to remain airborne are inhaled or settle on surfaces far from the bench. A splash shield does not contain them. Fix by listing every aerosol-generating step in the procedure, moving those steps into the cabinet, and documenting the change. If the cabinet is unavailable, the work waits. A backup plan that routes work to an open bench after hours defeats the control entirely.

### 3. Poor cabinet technique

A certified cabinet used badly offers less protection than an uncertified one used well. The common errors are placing items on the front grille, which disrupts the air curtain, blocking the rear grille with waste bags, working with the sash raised above the marked line, using a gas burner inside the cabinet, moving arms in and out repeatedly, and overloading the work surface.

Two related mistakes deserve their own mention. First, UV lamps are sometimes treated as a substitute for surface disinfection and for cleaning up after work. UV light penetrates poorly, is blocked by dust and shadow, and carries eye and skin hazard. Second, cabinet certification is often treated as permanent. Standard practice is annual certification by a qualified technician, plus recertification after any relocation or repair.

Safety note: never use a flammable gas burner inside a BSC. The airflow pattern disturbs the flame and the heat damages the filter. Use a disposable loop or a micro-incinerator instead.

### 4. PPE errors

The recurring errors are an open lab coat worn outside the lab, open-toe shoes in a wet lab, jewelry and watches worn under gloves, contact lenses worn without eye protection, and gloves worn for tasks that do not need them and then transferred to door handles, keyboards, and phones. Glove removal is a skill: pinch the outside of one glove at the wrist, peel it inside out, slide two fingers under the remaining glove at the wrist, and peel it off over the first, all without touching the skin.

Double gloving is not a substitute for technique. It is a control for specific high-risk tasks such as handling large volumes of infectious material or working with sharps under a defined protocol. Wearing two pairs for routine bench work reduces dexterity and increases the chance of a tear.

### 5. Hand hygiene and the exit sequence

Handwashing is the last line of defense and it fails in predictable ways: no soap at the sink, a sink used for storage, gloves removed but hands not washed, and the door handle touched with a gloved hand on the way out. The correct exit sequence is glove removal, handwash, lab coat removal at the door if the coat does not leave the lab, then exit. If a lab coat must leave the room, it travels in a closed bag and is laundered by a service that handles contaminated linen.

### 6. Spill response and disinfectant mistakes

Spill response errors cluster around three variables: concentration, contact time, and organic load. Household bleach at a 1:10 dilution is a standard working concentration for blood and body fluid spills, prepared fresh and used within 24 hours because available chlorine decays. A 1:100 dilution is sometimes used for low-organic-load surfaces but is inadequate for a protein-rich spill. Quaternary ammonium compounds are widely used but are inactivated by organic material and by some detergents. Alcohols are effective against many enveloped viruses and vegetative bacteria but are not sporicidal and evaporate before contact time is complete.

Safety note: never mix bleach with acid or ammonia products. Acid releases chlorine gas and ammonia forms chloramine vapors. Both are respiratory hazards in a confined space. Store bleach away from acids and label every secondary container with the concentration and the preparation date.

The other common spill error is technique. Staff spray disinfectant onto a spill and wipe outward, which spreads contamination. The correct order is to contain with absorbent material, apply disinfectant from the edge inward, allow the full contact time (typically 10 to 30 minutes depending on the product label), then remove with absorbent towels and clean the surface again.

### 7. Waste segregation and autoclave errors

The most frequent waste error is treating the autoclave as a formality. Indicator tape changing color shows that the load reached temperature. It does not show that steam penetrated the center of a dense bag. Validation requires biological indicators with defined spore species and a defined schedule, plus a log that records cycle number, temperature, pressure, time, and indicator result.

Other errors include overfilling bags so steam cannot penetrate, using bags that melt and seal the load, mixing sharps into soft waste, holding biohazardous waste in corridors for extended periods, and assuming that a bag marked with a biohazard symbol has been treated when it has only been staged. Standard autoclave conditions for biohazardous waste are 121°C at 15 psi for 15 to 30 minutes, with the actual time set by load type and validated locally.

### 8. Sharps handling

Recapping a needle is the classic error, and it persists because the one-handed scoop method is slower. Every recapping event is a percutaneous exposure opportunity. The fix is a written sharps procedure, immediate disposal into a puncture-resistant container at the point of use, containers replaced at three-quarters full, and a ban on hand-carrying open sharps between rooms.

Glass Pasteur pipettes, broken slides, and scalpel blades are equally common sources. Anyone handling broken glass needs forceps and a scoop, not gloved fingers, and the sharps container must be within arm's reach before the glass breaks.

### 9. Centrifuge and aerosol release

Centrifuges concentrate aerosols. A tube that cracks at speed releases material into the chamber, and opening the lid immediately after the rotor stops delivers that aerosol to the operator's face. Standard practice is to use sealed safety cups or rotors, inspect tubes and O-rings before each run, balance properly, and wait at least 10 minutes at BSL-2 and 30 minutes at BSL-3 before opening the lid. If a tube breaks, the lid stays closed for the full wait, and the rotor is opened inside the BSC.

### 10. Access, signage, and training records

BSL-2 requires restricted access when work is in progress, a biohazard sign at the entry that names the agent and the responsible person, and a biosafety manual available to all staff. The errors are a sign with the wrong agent, a sign that lists every agent in the building, no contact number that reaches a person after hours, and access control that amounts to an unlocked door with a printed notice.

Microbiology biosafety training records often show completion without competence. A useful record names the specific procedures the person was trained on, the trainer, the date, and the method of assessment. Annual refresher training should include a practical component such as a mock spill, a cabinet run, and a glove removal demonstration.

### 11. Equipment decontamination before service or transfer

A centrifuge, incubator, or computer that leaves a BSL-2 room carries contamination. Service engineers and movers are exposed when the equipment arrives without a decontamination certificate. The fix is a form that documents the agent handled, the method used (for example, formaldehyde gas for enclosed equipment or a validated wipe-down for surfaces), the date, and the signature of the person who performed it.

### 12. Personal items, food, phones, and cosmetics

Food and drink, gum, cigarettes, and cosmetics are prohibited in the laboratory. Mobile phones are the most common breach because they are used for timers, photography, and communication. A phone used with gloved hands becomes a contaminated object that then goes to the ear and the pocket. The practical fix is a designated phone location outside the work area and a clear rule about glove removal before any device handling.

### 13. Pest control and housekeeping

Rodents, insects, and standing water undermine every other control. Open waste, accumulated cardboard, and unsealed entry points create a route for agent dispersal. Housekeeping errors such as dry sweeping, which resuspends particles, and unlabeled secondary containers, which invite misuse, are easy to correct once they are on an inspection list.

### 14. Occupational health and immunization

BSL-2 programs are expected to offer hepatitis B vaccination to staff with exposure to blood or body fluids, provide baseline serum storage where indicated, and have a written exposure protocol that includes immediate first aid, reporting, and post-exposure evaluation. A common error is a protocol on file that no one has read since onboarding. Another is staff who do not know where the exposure reporting form is kept.

### 15. Inventory and oversight failures

The rarest but most consequential errors involve accountability. An agent that arrives without a log entry, a transfer that happens without a written record, a freezer inventory that no longer matches the contents, and a dual-use review that never occurred. These failures are rare in BSL-1 and BSL-2 work but they matter because a single unlogged transfer can undo years of good practice.

## How to Fix Biosafety Level 1 and 2 Mistakes: A Ten-Step Audit

A workable audit can be completed in one day by two people, one who knows the procedures and one who does not.

1. **List every procedure.** Include research, diagnostic, teaching, and maintenance tasks such as cleaning a centrifuge or changing a filter.
2. **Map each procedure to an agent and a concentration.** Include stocks, cultures, and clinical specimens.
3. **Mark aerosol-generating steps.** Vortexing, pipetting, opening tubes, loop work, sonication, and blending.
4. **Mark sharps steps.** Needles, blades, glass, and capillary tubes.
5. **Compare the mark to the control.** Cabinet, splash shield, closed system, or open bench.
6. **Walk the exit path.** Sink position, soap, towel, door handle, coat storage.
7. **Open every spill kit.** Check contents against the inventory and replace anything expired.
8. **Review sterilization records.** Confirm that biological indicators are run on schedule and that failures are documented and investigated.
9. **Observe three procedures live.** Do not announce the observation in advance. Record technique, not intention.
10. **Close each finding with a named owner and a date.** A finding without an owner is a note, not a fix.

The validation step deserves discipline. Lessons from experimental error analysis apply directly to biosafety verification: applying statistical judgment to an insufficient number of repeated measurements underestimates the real error, and the practical recommendation is to repeat a selected test under identical conditions five times and confirm that the results fall within the expected range [2]. Applied to a laboratory, this means that a single successful mock spill drill or one passing autoclave cycle does not validate the practice. Five repetitions with consistent results does more to prove that the control works when nobody is watching.

## Prevention Practices That Hold Up

### Build the risk assessment into intake

Every new assay, new agent, and new piece of equipment should trigger a short assessment form. The form should take under 30 minutes to complete. If the paperwork takes two hours, staff will skip it. Keep the form to one page and include a checkbox list for aerosol generation, sharps, splash, waste volume, and PPE.

### Train on observed competence

Microbiology biosafety training should be delivered at three points: before first independent work, after any procedure change, and annually as a refresher. Each point should include an observed practical demonstration. A trainer who has never watched a new hire remove gloves has not confirmed that the new hire can remove gloves.

### Protect institutional memory

The strongest programs do not depend on one experienced person staying forever. Institutional memory and strategic vigilance are recognized as a critical capacity in biosafety governance, and the loss of that memory is a predictable failure mode when staff turn over [3]. Practical steps include a manual with revision history, a training matrix that shows who is competent in which procedure, and a supervisor-level review of every assessment annually.

### Treat containment as a system, not a checklist

Modern containment workflows are validated as integrated systems that combine facility design, engineering controls, sample handling, transfer procedures, and decontamination protocols, and their reproducibility depends on the documentation of every element [4]. For a BSL-2 laboratory, the equivalent is a written workflow that ties the cabinet, the centrifuge, the waste path, and the spill procedure together so that a single change does not break the chain.

### Assign owners and dates to every control

Spill kits, eyewash stations, first aid kits, fire extinguishers, and sharps container replacement all need a named owner and a monthly check. Post the checklist at the station and record the initials and date. An unowned control decays.

### Review after every near miss

A near miss is free information. A cracked tube that stayed sealed, a glove tear noticed before contact, a spill that was contained by a tray. Each one should generate a short written review with a corrective action and a follow-up date.

### Schedule external review

An outside reviewer sees what familiarity hides. An annual review by someone from another department, or by a safety officer who does not work in the room, catches the shortcuts that the local team has stopped noticing.

## Common Mistakes and Limitations

- **Audits catch visible behavior, not aerosol generation.** A tidy bench can still produce a fine mist. Pair observation with procedure review.
- **Training completion is not competence.** A certificate proves attendance, not skill. Only observed performance proves skill.
- **A risk assessment is a snapshot.** It is accurate on the day it was written. Any change in agent, volume, or equipment invalidates it.
- **Engineering controls need maintenance.** An uncertified cabinet, a clogged filter, or a failed autoclave is a control that no longer exists.
- **Disinfectant labels are not interchangeable.** Concentration and contact time are product-specific. A disinfectant that works on one organism may fail on another.
- **BSL-1 is not the absence of practice.** Standard precautions still apply. Handwashing, decontamination, and sharps care are required even for agents not known to cause disease in healthy adults.
- **A single drill does not validate a procedure.** Repeat the drill and confirm consistent results before declaring the control effective.
- **Individual exposure decisions require professional judgment.** A physician or occupational health provider should evaluate any actual exposure, and a qualified biosafety professional should review the program itself.

## Frequently Asked Questions

### What is the most common biosafety mistake in a BSL-2 lab?

The most common mistake is a risk assessment that no longer matches the bench work. Staff then perform aerosol-generating steps on an open bench because the assessment never flagged them.

### Can I do BSL-2 work on an open bench?

Yes, for steps that do not generate aerosols or splashes. Any step that produces a mist, spray, or droplet cloud belongs in a biological safety cabinet or a validated closed system.

### How often should a biological safety cabinet be certified?

Once a year by a qualified technician, and again after any relocation, repair, or filter change. Certification is not a one-time event.

### Is 70% ethanol enough for every spill?

No. Ethanol is useful for many surfaces and enveloped viruses, but it is not sporicidal and it evaporates before a full contact time on a large spill. Use a product labeled for the organism and the organic load involved.

### How long should bleach stay in contact with a surface?

Follow the product label, which for a 1:10 household bleach dilution is typically 10 to 30 minutes. Prepare the dilution fresh and discard it after 24 hours.

### Do I need a biological indicator for every autoclave run?

Not every run, but on a defined schedule that covers each load type, plus after any repair or failed cycle. Indicator tape alone does not prove sterilization.

### How often should biosafety training be repeated?

Before first independent work, after any procedure change, and at least annually. Each cycle should include an observed practical demonstration.

### Can I take my lab coat or phone out of the lab?

A lab coat worn in the work area stays in the lab or travels in a closed bag to a designated laundry. A phone used with gloved hands should be treated as contaminated and never taken to the face.

<script type="application/ld+json">
{
  "@context": "https://schema.org",
  "@type": "FAQPage",
  "mainEntity": [
    {
      "@type": "Question",
      "name": "What is the most common biosafety mistake in a BSL-2 lab?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "The most common mistake is a risk assessment that no longer matches the bench work. Staff then perform aerosol-generating steps on an open bench because the assessment never flagged them."
      }
    },
    {
      "@type": "Question",
      "name": "Can I do BSL-2 work on an open bench?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Yes, for steps that do not generate aerosols or splashes. Any step that produces a mist, spray, or droplet cloud belongs in a biological safety cabinet or a validated closed system."
      }
    },
    {
      "@type": "Question",
      "name": "How often should a biological safety cabinet be certified?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Once a year by a qualified technician, and again after any relocation, repair, or filter change. Certification is not a one-time event."
      }
    },
    {
      "@type": "Question",
      "name": "Is 70% ethanol enough for every spill?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "No. Ethanol is useful for many surfaces and enveloped viruses, but it is not sporicidal and it evaporates before a full contact time on a large spill. Use a product labeled for the organism and the organic load involved."
      }
    },
    {
      "@type": "Question",
      "name": "How long should bleach stay in contact with a surface?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Follow the product label, which for a 1:10 household bleach dilution is typically 10 to 30 minutes. Prepare the dilution fresh and discard it after 24 hours."
      }
    },
    {
      "@type": "Question",
      "name": "Do I need a biological indicator for every autoclave run?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Not every run, but on a defined schedule that covers each load type, plus after any repair or failed cycle. Indicator tape alone does not prove sterilization."
      }
    },
    {
      "@type": "Question",
      "name": "How often should biosafety training be repeated?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Before first independent work, after any procedure change, and at least annually. Each cycle should include an observed practical demonstration."
      }
    },
    {
      "@type": "Question",
      "name": "Can I take my lab coat or phone out of the lab?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "A lab coat worn in the work area stays in the lab or travels in a closed bag to a designated laundry. A phone used with gloved hands should be treated as contaminated and never taken to the face."
      }
    }
  ]
}
</script>

## Further Reading

- [[Sources of error and risks in CT based navigation].](https://pubmed.ncbi.nlm.nih.gov/12056279/)
- [Errors and mistakes in the ultrasound diagnostics of the liver, gallbladder and bile ducts.](https://pubmed.ncbi.nlm.nih.gov/26673428/)

## Sources

1. [Biosafety considerations and risk reduction strategy for a new veterinary faculty building and teaching hospital in Sweden.](https://pubmed.ncbi.nlm.nih.gov/32864050/)
2. [Tutorial review of error evaluation in experimental water research at the example of membrane filtration.](https://pubmed.ncbi.nlm.nih.gov/38579944/)
3. [Global health security and smallpox: A strategic review of post-eradication biosafety and preparedness.](https://pubmed.ncbi.nlm.nih.gov/42978405/)
4. [Biosafety level-3 cryo-electron tomography of active severe acute respiratory syndrome coronavirus 2 virions.](https://pubmed.ncbi.nlm.nih.gov/42989482/)