# Autoclave Bags and Sharps Containers: How to Choose

An autoclave bag is a heat-tolerant, steam-permeable bag that holds soft, non-sharp waste inside a steam sterilizer. A sharps container is a rigid, puncture-resistant, closable box that holds needles, blades, broken glass, and other items that can cut through plastic film. The rule of thumb that settles almost every case: if the object could puncture a bag and cut the person who lifts it, it goes in a rigid sharps container. If it is soft, non-sharp, and destined for steam decontamination, it goes in an autoclave bag. Nothing goes in both.

That single distinction drives the rest of this guide. The two products look like generic disposables on a supply order, and that is where most purchasing mistakes begin. A bag chosen for the wrong polymer melts onto the autoclave shelf and ruins the load. A sharps container chosen for the wrong size gets overfilled in three days and becomes a hand-injury hazard. Getting this right is one of the small, unglamorous skills that separates a lab that runs cleanly from one that generates incident reports. It is also the sort of concrete detail that a good microbiology course drills early, because waste handling sits at the intersection of biosafety, sterilization science, and occupational law.

## Autoclave Bags and Sharps Containers in Microbiology: The Basic Split

Microbiology generates two physically different kinds of contaminated material. The first is bulky, soft, and wet: agar plates, culture flasks, pipette tips, gloves, gowns, absorbent bench paper, filtration membranes, and plastic labware. Most of it can be rendered non-infectious by steam under pressure, which denatures proteins and kills even bacterial endospores when the cycle is properly validated. The second kind is sharp and rigid: hypodermic needles, scalpels, razor blades, broken glass slides, Pasteur pipettes, capillary tubes, and contaminated broken ampoules. These cannot be reliably sterilized in a thin bag, because the sharp edge punctures the film from the inside during handling, loading, or compaction.

The two streams therefore need two different engineering solutions. One solution prioritizes steam access. The other prioritizes containment and puncture resistance. A product designed for one job usually fails at the other.

### A four-question decision list

Use this sequence at the bench. It takes about five seconds and it resolves nearly every ambiguous item.

1. Can the item cut or puncture? If yes, it goes in a sharps container, full stop, regardless of whether it is also contaminated.
2. Is the item soft, non-sharp, and intended for steam treatment in-house? If yes, it goes in a steam-rated autoclave bag.
3. Is the item a liquid? Liquids need vented, rigid polypropylene containers or validated liquid cycles, not sealed bags.
4. Is the item going off-site for commercial treatment? Follow the transport packaging rules for regulated medical waste, which are stricter than in-house handling rules.

Anything that fails all four questions belongs in the general biohazard waste stream defined by your institution's exposure control plan, not in a bag or a sharps box by default.

## Side-by-Side Comparison

This table is the working core of the article. Read it as a specification sheet rather than a summary.

| Attribute | Autoclave bag | Sharps container |
|--|--|--|
| Primary function | Hold non-sharp waste for steam sterilization | Hold sharp waste for containment and disposal |
| Typical polymer | Polypropylene, sometimes a polypropylene and polyamide blend | Rigid polypropylene, often with recycled content |
| Heat tolerance | Rated for steam cycles, commonly up to about 121 to 135 C (250 to 275 F) | Not designed for steam. Deforms well below sterilization temperature |
| Steam permeability | Yes, by design. Steam must pass through the film | No. The wall is a barrier |
| Puncture resistance | Low. A needle passes through easily | High. The wall is tested for puncture and drop performance |
| Closure | Twist and tie, tape, or self-sealing adhesive strip, closed but not airtight | Snap lid, gasketed lid, or locking final closure with audible engagement |
| Labeling | Printed biohazard symbol, often with a process indicator stripe | Biohazard symbol, the word "SHARPS," permanent molded markings |
| Fill limit | About two-thirds full, so the bag can be gathered and closed | Stop at the visible fill line, typically around 75 percent capacity |
| Reuse | Single use | Single use. Never emptied or refilled |
| Verification | Class 1 process indicator on the bag, biological indicator in the load | None. The container is not a sterilizer, so there is nothing to verify |
| Failure mode | Melts, bursts, traps air, or wets out the load | Overfills, splits at a seam, or opens in transit |
| Typical setting | Microbiology lab, teaching lab, research bench, veterinary clinic | Any setting where injections, venipuncture, or sharps handling occurs |

Two rows deserve emphasis. The heat tolerance row explains why an autoclave bag and a sharps container are not interchangeable under any circumstances. The verification row explains why the two products carry different quality-control burdens. A bag is part of a sterilization process that must be monitored. A sharps container is part of a containment process that must be inspected.

## Autoclave Bags on Their Own Terms

An autoclave bag has four jobs at once. It must let steam in, keep waste in, survive the cycle without melting, and show that the cycle ran. Those four requirements explain nearly every design feature on the shelf.

### Steam has to reach the waste

Steam sterilization works by direct contact between saturated steam and the contaminated surface. A sealed bag blocks that contact and traps air, which is why an airtight bag will produce a failed cycle even at the correct temperature and pressure. Bags are made from thin polypropylene film, typically about 2 to 3 mil thick, because the film is permeable enough to let steam through while remaining strong enough to hold the load.

The practical loading rules follow from that physics:

- Fill the bag to about two-thirds capacity. A bag that is stuffed cannot be gathered, and trapped air pockets form in the center of the mass.
- Leave the bag open or loosely gathered during the cycle. Many laboratories fold the top over rather than sealing it.
- Do not double-bag. Two layers of film cut steam penetration and slow the drying phase.
- For dry loads such as wrapped glassware or empty plasticware, add a small volume of water, roughly 50 to 100 mL, so the load generates its own steam.
- Stand bags upright in a polypropylene tray so that condensate does not pool around the load.
- Do not pack bags tightly against the chamber wall or against each other. Steam needs a path.

### Polymer choice decides the cycle

Polypropylene is the standard polymer for steam autoclave bags because its softening point sits comfortably above the usual cycle temperatures. Standard steam cycles run at 121 C (250 F) with about 15 psi of pressure above atmospheric for 15 to 30 minutes in a gravity displacement sterilizer, or at 132 C (270 F) with about 27 psi for roughly 4 minutes, or at 134 C in a pre-vacuum cycle for 3 to 4 minutes. A bag rated to 135 C covers all of them. Cheap bags rated only to 121 C will deform in a 134 C cycle.

Polyethylene is the polymer behind many red biohazard transport bags, and it is the wrong choice for an autoclave. Polyethylene softens well below the 121 C cycle temperature, so a polyethylene bag will slump, fuse to itself, and coat the chamber shelf. Nylon and polyethylene are used for ethylene oxide bags because that process runs at much lower temperatures, typically 37 to 55 C. A bag rated for ethylene oxide is not a steam bag.

Compostable and biodegradable bags made from starch blends or polylactic acid are designed for ambient composting. They are not validated for a 121 C steam cycle and should not be substituted on the assumption that a "greener" bag is a safe bag.

### Bag formats

The catalog offers more variety than most people expect. The common formats and their uses:

- **Flat bags with an indicator stripe.** The most common laboratory format. The printed Class 1 process indicator changes color when exposed to steam, which confirms the bag was in a cycle, not that the load was sterilized.
- **Gusseted bags.** Expand to hold bulky items such as discarded gowns or large flask assemblies.
- **Self-sealing bags with an adhesive strip.** Convenient for transport between rooms, but the seal must be broken or left loose before autoclaving.
- **Bags with a tie or twist closure.** Simple and reliable when gathered loosely.
- **High-temperature bags.** Rated for 135 C or above and required if the facility runs a 134 C pre-vacuum cycle.
- **Heavy-duty bags.** Thicker film for loads containing rigid plasticware with rough edges.

For sterile barrier packaging of instruments and devices, the relevant standard is ISO 11607, which covers materials, sterile barrier systems, and the validation of forming and sealing processes. Waste bags are not sterile barrier systems and are not validated under that standard. They are process consumables whose performance is verified by the load monitoring program, not by the bag manufacturer alone. That is a meaningful distinction for anyone writing a laboratory quality manual.

### Verifying the bag and the cycle

A color change on the bag stripe is a Class 1 process indicator under ISO 11140. It tells you the item saw the conditions of a cycle. It does not tell you the load was sterile. For that, you need a biological indicator and, for complex or high-value loads, physical validation.

For steam, the reference biological indicator organism is *Geobacillus stearothermophilus*, historically known as *Bacillus stearothermophilus*, commonly cited as ATCC 7953. Its spores are substantially more heat resistant than the vegetative bacteria and viruses the cycle is meant to destroy, so survival of the indicator signals a failed cycle. ANSI/AAMI ST79, the comprehensive guide for steam sterilization in health care facilities, sets the monitoring framework widely adopted in clinical labs, including at least weekly biological indicator testing and biological indicator testing with every load containing implants.

Place the biological indicator where steam is least likely to reach. In a bag load, that is usually the center of the densest, most tightly packed bag near the front or the bottom of the chamber, depending on the sterilizer's documented cold spot. Running the indicator in an empty chamber proves nothing about the load.

For waste decontamination, many facilities run longer cycles than they would for instruments, often 30 to 60 minutes at 121 C, because the loads are bulky, heterogenous, and hard to penetrate. Those times must be established by validation for the specific load configuration, not copied from a wall chart.

## Sharps Containers on Their Own Terms

A sharps container is an engineering control. In the United States, its required features come from the OSHA Bloodborne Pathogens Standard, 29 CFR 1910.1030, which was strengthened by the Needlestick Safety and Prevention Act of 2000. The standard requires containers that are closable, puncture resistant, labeled or color-coded, and leakproof on the sides and bottom. It also requires that containers be replaced routinely and not allowed to overfill, and that contaminated sharps not be bent, recapped, or removed unless the employer can demonstrate that no alternative is feasible or that a recapping must be done by a mechanical device or a single-handed technique.

Those five words, closable, puncture resistant, labeled, leakproof, and not overfilled, are the entire specification in miniature. Every purchasing decision should map onto one of them.

### Materials and construction

Rigid polypropylene is the standard material because it holds its shape under load, resists puncturing from the inside, and can be molded with integrated features such as needle disengagement ports and lid hinges. Manufacturers test containers for puncture resistance and drop performance, and a container that has cracked in transit or been crushed in storage should be discarded rather than put into service.

Wall thickness varies with container volume. Small countertop units are thinner. Large floor-standing units carry more weight and are built accordingly. A container that flexes when you press the side is not one to trust with a full load of needles.

### Geometry, lids, and closure

Sharps containers come in several configurations, and the right one depends on where the container sits and who uses it.

- **Countertop and bracket-mounted units.** Small, 1-quart to 2-gallon sizes, placed at the point of use. Wall brackets keep the opening at a comfortable height.
- **Floor-standing units with a dolly or cart.** Larger volumes for wards, phlebotomy stations, and procedure rooms.
- **Drop-in or screw-cap lids.** Common in small point-of-use containers.
- **Tortuous-path lids.** The opening is offset so that a filled container cannot easily spill its contents if tipped. This is a meaningful safety feature in any setting where the container might be knocked over.
- **Temporary and final closure.** Most containers have a temporary closure for use during filling and a final locking closure that engages permanently for disposal, often with an audible click or a visible lock indicator.

The final closure matters because it defines the boundary between a container in service and a package in transport. Once final closure is engaged, the container should not be reopened.

### Placement and fill management

Placement decides whether staff actually use the container. The working standard is to put a sharps container within arm's reach of the point where the sharp is generated, at a height that does not require reaching overhead or bending to the floor. A container in a distant utility room will be bypassed, and a bypassed container produces recapping, which is exactly the behavior the standard is written to prevent.

Fill management is equally practical. Most containers carry a visible fill line near 75 percent capacity. Close and replace the container before contents reach that line. Never shake a container to settle contents, never press contents down, and never transfer contents from one container to another. If a sharp is lodged in the opening, use forceps rather than fingers.

### Labeling and transport

The container must display the biohazard symbol or be color-coded in fluorescent orange or orange-red with contrasting lettering under the OSHA standard, and the word "SHARPS" should be visible and permanent. In-house containers and containers leaving the building are governed by different rules. Once a container leaves the facility, it becomes a package under Department of Transportation hazardous materials regulations, and regulated medical waste is shipped under the UN 3291 entry. The outer packaging requirements are stricter than anything printed on a countertop container.

If a container leaks or is damaged, it goes into a secondary container that meets the same requirements. A leaking sharps container is an exposure incident waiting to happen, and it should be treated as one.

### Autoclavable sharps containers

A narrow category exists for containers that are designed to be processed in a sterilizer, usually paired with a sterilizer-grinder system that both decontaminates and renders the sharps unrecognizable. These containers are labeled for that use and validated for it. A standard sharps container placed in an autoclave will deform, split, and spill. It is not an autoclave bag with a lid.

## The Criteria That Decide Real Choices

Once you understand both products, the comparison narrows to five practical criteria. The first two eliminate most of the catalog immediately.

### Criterion 1: Sterilization method

Match the bag to the process, not to the catalog category. Steam requires polypropylene rated above the cycle temperature. Ethylene oxide requires a compatible film such as nylon or polyethylene. Vaporized hydrogen peroxide and other low-temperature processes have their own material compatibility lists. If your facility runs more than one process, buy separate, clearly labeled stock and store it in separate drawers.

For sharps containers, the only relevant question is whether the container will be treated in-house by a validated sterilizer-grinder system. If not, the container is a containment device and the material question is settled.

### Criterion 2: Waste stream and regulatory destination

In-house decontamination and off-site commercial treatment are different supply chains. Sterile barrier packaging, waste bags, transport bags, and shipping containers are each governed by different rules, and buying a product outside its intended destination creates a compliance gap that shows up in an inspection.

### Criterion 3: Volume and throughput

A large research lab may fill multiple gusseted bags a day and generate a steady trickle of sharps. A small teaching lab may generate one bag per week and almost no sharps. Size the container to the generation rate, not to the storage shelf. Oversized floor containers in low-volume areas sit half-empty for months, which is both a cost problem and a housekeeping problem. Undersized countertop containers in high-volume areas get overfilled, which is an injury problem.

### Criterion 4: Physical space and workflow

An autoclave bag needs to fit the sterilizer tray and leave a steam path. A sharps container needs a bracket, a shelf, or a floor footprint at the point of generation. Measure the space before ordering. A container that does not fit the available bracket will end up on the floor, and a floor container gets bypassed.

### Criterion 5: Consequence of failure

Cost per unit is the wrong headline number for both products. The failure modes are asymmetric. A failed bag cycle means a re-run, a contaminated load, and possibly a chamber cleanup. A failed sharps container means a needlestick injury, an exposure workup, and a recordable incident. Weight the decision toward the failure that harms a person.

## Worked Scenarios

### Scenario: A small teaching laboratory

A university microbiology teaching lab runs two benchtop autoclaves, generates about twenty agar plates per session, and has no sharps generation beyond the occasional broken slide. The right kit is a stock of 2 to 3 mil polypropylene flat bags with an indicator stripe, rated to at least 135 C, sized so that a stack of five plates fits in a two-thirds fill. A small 1-quart countertop sharps container on a wall bracket handles the broken slides. Biological indicators are run weekly by the lab manager. Nothing in this scenario justifies a floor-standing sharps unit or an ethylene oxide bag.

### Scenario: A clinical microbiology bench

A hospital bench processes sputum, blood cultures, and wound swabs, and it also receives contaminated needles from blood culture collection. Two streams exist side by side. Autoclave bags hold plates, tubes, and disposable loops. A wall-mounted 2-gallon sharps container at the bench handles the needles. The container has a tortuous-path lid because the bench is crowded and a tipped container would be a disaster. The bags run in a validated 121 C cycle with a demonstrated cold spot in the center of the load.

### Scenario: A BSL-2 research laboratory

A research lab works with bacterial cultures at [biosafety level](/knowledge/diagnostics/emerging-tech/biosafety-levels-1-4-a-comparative-framework-for-diagnostic-laboratory-design) 2 under the framework described in the *Biosafety in Microbiological and Biomedical Laboratories* manual published by CDC and NIH, and in the WHO *Laboratory [Biosafety Manual](/knowledge/diagnostics/emerging-tech/biosafety-manual-development-a-template-for-diagnostic-laboratories)*, fourth edition. All waste is decontaminated before leaving the room. The lab uses double-door pass-through autoclaves where available, with bags loaded inside the containment area and unloaded outside. Sharps go into containers that are closed inside the room before removal. The critical supply decision here is not brand but compatibility: the bag must survive the facility's validated cycle, and the container must fit the pass-through port.

### Scenario: A veterinary clinic

A small animal clinic generates used hypodermic needles, scalpel blades, and contaminated gauze. The gauze is soft and non-sharp but is usually not autoclaved on site, so it goes into a red biohazard bag for commercial pickup rather than an autoclave bag. The needles and blades go into a 1-gallon countertop sharps container at the treatment table, with a second container in the surgery suite. The clinic's autoclave is used for instrument packs, which require ISO 11607 compliant sterile barrier packaging, not waste bags. Mixing these functions is one of the most common errors in small-practice settings.

## Autoclave Bags and Sharps Containers: Pros and Cons

The pros and cons are easiest to see when each product is judged against its own purpose rather than against the other.

**Autoclave bags, pros.** Low cost per load. Steam permeable, so the sterilization process works as designed. Available with process indicators that document the cycle. Gusseted and heavy-duty formats cover bulky and rough-edged loads. Compact storage before use. Compatible with pass-through autoclave workflows.

**Autoclave bags, cons.** No puncture resistance. Fail visibly and messily when the wrong polymer is chosen. Require correct loading, which means training. Cannot be sealed airtight, which creates transport handling issues. Provide no containment if the bag bursts in the chamber. Indicator color change is often mistaken for proof of sterilization.

**Sharps containers, pros.** Rigid puncture-resistant containment. Closable with a permanent final lock. Labeled and color-coded to the OSHA specification. Available in sizes and mounting options that fit the point of generation. Audible and visible closure confirmation. Designed to be handled without opening.

**Sharps containers, cons.** Not sterilizable in a standard autoclave. Bulkier to store and ship than bags. Require bracket or floor space. Fill management depends on staff attention. Disposal cost is higher than for autoclaved waste. Cannot absorb liquids, so a container that receives fluid will leak.

The honest summary is that these products are complements, not competitors. A lab needs both, and the supply decision should be made for each stream separately.

## Common Mistakes and Limitations

**Using a red transport bag in the autoclave.** Red polyethylene biohazard bags are built for collection and transport, not for steam. They soften and fuse in a 121 C cycle. If you want a red bag that can be autoclaved, verify the polymer and the temperature rating on the label rather than the color.

**Sealing the bag airtight before loading.** A sealed bag traps air, blocks steam, and produces a failed cycle. Gather the top loosely or fold it over.

**Judging sterilization by the indicator stripe.** A Class 1 indicator confirms exposure to process conditions. It does not confirm sterility. Biological indicators and physical monitoring carry that burden.

**Skipping the cold spot.** A biological indicator placed on top of the load will pass while the center fails.

**Putting needles in an autoclave bag.** The needle punctures the film, the bag leaks contaminated material into the chamber, and the person who handles the bag gets exposed. This is the single most consequential mix-up in the category.

**Overfilling a sharps container.** OSHA requires containers to be replaced routinely and not allowed to overfill. The practical rule is to close before the fill line. Contents above the line protrude into the lid mechanism and cause needlesticks during closure.

**Recapping by hand.** The standard restricts recapping to cases where no alternative is feasible or where a mechanical device or single-handed technique is used. A recapping device at the point of generation removes the temptation.

**Using a sharps container as a general biohazard bin.** Gauze, gloves, and wrappers belong in the biohazard waste stream, not in a sharps container. Bulky non-sharp items fill the container quickly and increase the chance of a sharp protruding through the lid.

**Ignoring the exposure control plan.** The OSHA standard requires a written exposure control plan, engineering controls, and a sharps injury log for employers with eleven or more employees. The log is not paperwork for its own sake. It is how a facility learns which container or device is failing.

**Treating the bag as a sterile barrier.** Waste bags are process consumables. Instrument packaging is a sterile barrier system governed by ISO 11607. These are different regulatory categories and different purchasing decisions.

**Assuming a container is autoclavable because it is polypropylene.** Many sharps containers are polypropylene, but most are not designed or validated for a steam cycle. Only use a container labeled for that purpose with a validated system.

**Limitations of the products themselves.** A bag cannot contain a puncture. A sharps container cannot sterilize. Neither product substitutes for a validated cycle, a trained operator, or a written procedure. Individual facility requirements vary with the sterilizer, the load, and the waste destination, and a qualified biosafety officer or sterilization supervisor should sign off on the final configuration.

## Frequently Asked Questions

### Can needles go in an autoclave bag?

No. A needle punctures the film, contaminates the chamber, and exposes whoever handles the bag. Needles and other sharps belong in a rigid, closable, puncture-resistant sharps container.

### Can sharps containers be autoclaved?

Generally no. Standard sharps containers deform and split at steam sterilization temperatures. The exception is a container specifically labeled and validated for use with a sterilizer-grinder system.

### Are red bags the same as autoclave bags?

No. Red biohazard bags are usually polyethylene and are made for collection and transport. Autoclave bags are polypropylene and must be rated above the cycle temperature. Color does not indicate polymer, so check the label.

### How full should a sharps container be?

Close and replace it before contents reach the visible fill line, which is typically around 75 percent of capacity. Never shake, compress, or transfer the contents.

### Do autoclave bags need to be sealed before the cycle?

They need to be closed enough to hold the waste, but not airtight. Steam must pass through the film, so the top is usually gathered loosely or folded over.

### What is the difference between sterilization and decontamination?

Sterilization destroys all microorganisms including bacterial endospores. Decontamination reduces contamination to a level considered safe for handling and disposal. Both can use the same autoclave, but they use different validated cycles and different monitoring.

<script type="application/ld+json">
{
  "@context": "https://schema.org",
  "@type": "FAQPage",
  "mainEntity": [
    {
      "@type": "Question",
      "name": "Can needles go in an autoclave bag?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "No. A needle punctures the film, contaminates the chamber, and exposes whoever handles the bag. Needles and other sharps belong in a rigid, closable, puncture-resistant sharps container."
      }
    },
    {
      "@type": "Question",
      "name": "Can sharps containers be autoclaved?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Generally no. Standard sharps containers deform and split at steam sterilization temperatures. The exception is a container specifically labeled and validated for use with a sterilizer-grinder system."
      }
    },
    {
      "@type": "Question",
      "name": "Are red bags the same as autoclave bags?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "No. Red biohazard bags are usually polyethylene and are made for collection and transport. Autoclave bags are polypropylene and must be rated above the cycle temperature. Color does not indicate polymer, so check the label."
      }
    },
    {
      "@type": "Question",
      "name": "How full should a sharps container be?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Close and replace it before contents reach the visible fill line, which is typically around 75 percent of capacity. Never shake, compress, or transfer the contents."
      }
    },
    {
      "@type": "Question",
      "name": "Do autoclave bags need to be sealed before the cycle?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "They need to be closed enough to hold the waste, but not airtight. Steam must pass through the film, so the top is usually gathered loosely or folded over."
      }
    },
    {
      "@type": "Question",
      "name": "What is the difference between sterilization and decontamination?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Sterilization destroys all microorganisms including bacterial endospores. Decontamination reduces contamination to a level considered safe for handling and disposal. Both can use the same autoclave, but they use different validated cycles and different monitoring."
      }
    }
  ]
}
</script>

## Further Reading

- [Clear aligner interface properties, oral microbial ecology, and strategies for biofilm control and caries-risk mitigation: a scoping review.](https://pubmed.ncbi.nlm.nih.gov/42997201/)
- [Integrating microbiome source tracking, machine learning, and precision intervention for data-driven control of microbial contamination in food processing.](https://pubmed.ncbi.nlm.nih.gov/42977414/)
- [A triple-responsive smart controlled-release system for antifungal volatiles: toward sustainable agriculture via pathogen-targeted control and crop growth promotion.](https://pubmed.ncbi.nlm.nih.gov/42983820/)
- [Engineering functional microbial inoculants for cigar tobacco fermentation: metabolic mechanisms, synthetic consortia, and translational challenges.](https://pubmed.ncbi.nlm.nih.gov/42967955/)
- [Antimicrobial-resistant Salmonella and Escherichia coli in fresh vegetables from Ouagadougou, Burkina Faso: Prevalence, β-lactamase genes, and food safety implications.](https://pubmed.ncbi.nlm.nih.gov/42986907/)
- [Long-term safety and efficacy of twice-yearly depemokimab in type 2 asthma with an eosinophilic phenotype: 2-year integrated analysis of the phase 3a SWIFT-1 and SWIFT-2 studies and AGILE open-label extension.](https://pubmed.ncbi.nlm.nih.gov/42956801/)