# DNA Dilution Calculator: From Nanodrop Readings to PCR Templates

By the end of this guide, you will be able to take a raw DNA concentration reading from a Nanodrop or fluorometer, calculate exactly how much stock and diluent to mix, prepare a PCR template at the concentration your assay expects, and verify that the dilution actually worked. You need only three inputs on hand: your stock concentration in ng/µL, your target concentration, and your final volume. A calculator, spreadsheet, or ten lines of Python does the rest.

Dilution sounds trivial, and the arithmetic is. What makes dilutions fail in real labs is everything around the arithmetic: concentrations measured on the wrong instrument, stock volumes too small to pipette accurately, inhibitors carried into the reaction, and templates so degraded that the dilution factor stops mattering. This article walks through the calculation, the judgment calls around it, and the mistakes that cost people the most time.

## What a DNA Dilution Calculator Actually Computes

A DNA dilution calculator is a small biology calculator built on one equation, the dilution formula:

**C1 × V1 = C2 × V2**

Here C1 is your stock concentration, V1 is the volume of stock you need, C2 is your target concentration, and V2 is your final volume. Every dilution calculator, whether a web app, a spreadsheet, or a notebook script, solves this equation for V1 and reports how much diluent to add.

The same engine powers a broader family of tools that bioinformatics calculators and bench calculators share: dilution factor, [serial dilution](/knowledge/molecular-biology/serial-dilution) chains, molarity conversions, and copy number per microliter. The dilution step sits between quantification and amplification, and errors there propagate into every downstream result.

A useful summary of the core formulas:

| Quantity | Formula | Worked example |
|--|--|--|
| Stock volume needed | V1 = (C2 × V2) ÷ C1 | 5 ng/µL × 100 µL ÷ 85 ng/µL = 5.88 µL |
| Diluent volume | V2 - V1 | 100 - 5.88 = 94.12 µL |
| Dilution factor | DF = C1 ÷ C2 | 85 ÷ 5 = 17-fold |
| Multi-step dilution | Cfinal = C1 ÷ DF^n (n steps) | 85 ng/µL ÷ 10^3 = 0.085 ng/µL after three 1:10 steps |
| Copy number (dsDNA) | copies/µL = (ng/µL × 6.022×10^23) ÷ (660 × length in bp) | 5 ng/µL of a 500 bp fragment = 9.1×10^9 copies/µL |

The dilution formula is textbook chemistry. The copy number formula uses 660 g/mol as the average mass of one base pair of double-stranded DNA, a standard approximation in [molecular biology](/blog/careers/molecular-biology).

## DNA Dilution Calculator Step by Step

This walkthrough starts from a Nanodrop reading and ends with a PCR-ready template. Suppose your extraction gave you a stock reading of 85 ng/µL, and your polymerase manufacturer or lab protocol calls for roughly 5 ng/µL of template in a reaction you want to assemble from a 100 µL working stock.

### Step 1: Record an honest stock concentration

Measure your stock on a Nanodrop or equivalent spectrophotometer, or better, on a fluorometer such as a Qubit with a dsDNA-specific dye. Record the value in ng/µL and note the purity ratios at the same time. A Nanodrop reports absorbance at 260 nm and converts it to concentration using the convention that 1 absorbance unit corresponds to about 50 µg/mL of double-stranded DNA. Pure DNA typically shows a 260/280 ratio near 1.8 and a 260/230 ratio near 2.0 or higher. Ratios far from these values warn you that protein, phenol, salt, or RNA may be inflating the reading, which matters because you are about to trust this number in your arithmetic.

### Step 2: Define the target and final volume

Write down two numbers before touching the calculator. Here, C2 = 5 ng/µL and V2 = 100 µL. Skipping this step is the most common way people produce a dilution that is mathematically correct but useless, for example 40 µL of a 5 ng/µL stock when the protocol needed enough working solution for a dozen reactions.

### Step 3: Run the calculation

![Worked calculation: 5.88 uL of 85 ng/uL stock plus 94.12 uL of diluent makes 100 uL at 5 ng/uL.](/article-images/illustrations/dna-dilution-calculator-stock-volume-calc.webp)

In Python:

```python
def dilution(c_stock, c_target, v_final):
    v_stock = (c_target * v_final) / c_stock
    v_diluent = v_final - v_stock
    return v_stock, v_diluent

v_stock, v_diluent = dilution(c_stock=85.0, c_target=5.0, v_final=100.0)
print(f"Stock volume:   {v_stock:.2f} uL")
print(f"Diluent volume: {v_diluent:.2f} uL")
print(f"Dilution factor: {85.0 / 5.0:.1f}x")
```

Expected output:

```text
Stock volume:   5.88 uL
Diluent volume: 94.12 uL
Dilution factor: 17.0x
```

In a spreadsheet, the same result comes from `=(5*100)/85` in one cell and `=100-B1` in the next, where B1 holds the stock volume. Any web-based dna dilution calculator performs the identical operation, so the tool you choose matters far less than the inputs you give it.

### Step 4: Check pipetting feasibility

![Path checking whether the calculated stock volume is too small to pipette, and moving to a two-step or serial dilution when it is.](/article-images/illustrations/dna-dilution-calculator-pipetting-feasibility.webp)

The calculation says to pipette 5.88 µL of stock. That is comfortable for a P20 or P200 set correctly. If the answer had been 0.6 µL, single-step pipetting would be unreliable, and you would move to a two-step or serial dilution instead. A practical rule from standard laboratory practice is to stay within the reliable middle of your pipette's range and avoid dispensing volumes near the bottom of what the tip can accurately deliver. When the required stock volume is too small, dilute in two stages rather than forcing one heroic pipette stroke.

### Step 5: Choose the diluent

For DNA intended as a PCR template, the standard choices are nuclease-free water or low-EDTA TE buffer (typically 0.1× TE, meaning 1 mM Tris and 0.1 mM EDTA). Water is convenient and leaves nothing behind. Low-salt TE buffers the DNA and protects it from nuclease damage and freeze-thaw adsorption, which matters if the working stock will sit in the freezer for weeks. High EDTA concentrations inhibit magnesium-dependent polymerases, which is why concentrated TE (10 mM Tris, 1 mM EDTA) is a storage buffer, not a reaction buffer.

### Step 6: Mix, label, and log

Add the diluent first, then the stock, so the pipette tip lands in liquid rather than onto a dry tube wall. Mix by gentle flicking or brief low-speed vortexing, then a quick spin. Label the tube with the calculated concentration, the date, the dilution factor, and the diluent. Write the entry into your [lab notebook](/knowledge/molecular-biology/lab-notebook) or electronic record at the bench, not afterward.

### Step 7: Verify

Re-measure the diluted working stock, ideally on a fluorometer, since fluorescence assays are more specific for double-stranded DNA than absorbance at low concentrations. The measured value should land close to 5 ng/µL. If it does not, resist the urge to blame the calculator, since the arithmetic almost never fails. The likely culprits are an inaccurate starting concentration or pipetting error, both covered below.

## From Nanodrop Readings to PCR Templates

The journey from a spectrophotometer reading to a working PCR template involves three separate conversions, and each one can go wrong independently.

### Conversion 1: absorbance to concentration

A Nanodrop infers concentration from light absorption at 260 nm. It cannot distinguish DNA from RNA, free nucleotides, or anything else that absorbs at that wavelength. This is why fluorometric quantification with a dsDNA-specific dye is preferred when template amount matters, for example in quantitative PCR or next-generation sequencing library preparation. Absorbance-based readings also lose accuracy at low concentrations, which is exactly the range where dilute PCR templates live.

### Conversion 2: concentration to copy number

Many modern assays, including digital PCR and CRISPR-based biosensors, are specified in copies per microliter rather than ng/µL. One published PCR-activated Cas12a biosensor for the carbapenem resistance gene blaKPC-2 showed a concentration-dependent response across roughly 10 to 10^7 copies per µL, with the signal compressing at the top of that range [1]. Converting mass concentration to copy number uses the formula in the table above:

```python
AVOGADRO = 6.022e23
BP_MASS = 660  # grams per mole per base pair, double-stranded DNA

def copies_per_ul(ng_per_ul, length_bp):
    mass_g = ng_per_ul * 1e-9
    return (mass_g * AVOGADRO) / (BP_MASS * length_bp)

print(f"{copies_per_ul(5.0, 500):.2e}")
```

Output:

```text
9.12e+09
```

So a 5 ng/µL working stock of a 500 bp fragment carries about 9 billion template molecules per microliter, far above what a typical PCR needs. This is why dilution, not concentration, is the daily reality of template preparation.

### Conversion 3: concentration to cycle threshold expectations

Once you know the template concentration, you can predict qPCR behavior. With an ideal assay, each 10-fold dilution shifts the cycle threshold (Ct, the PCR cycle at which fluorescence crosses the detection threshold) upward by about 3.3 cycles, because 10-fold fewer starting molecules need roughly log2(10), or 3.32, additional doubling cycles. Real assays deviate from this, but large deviations flag problems such as inhibition or pipetting error.

Dilution also interacts with template quality. In a study of vacuum-dried DNA stored under accelerated aging at 70 °C, samples heated for 0 to 30 hours showed no significant difference in Ct values compared with unheated controls, even after 1000-fold dilution, and a significant Ct increase appeared only under the harshest condition, 40 hours of heating combined with 1000-fold dilution [2]. That result carries a practical lesson: a well-preserved template tolerates aggressive dilution, while a marginal one does not, and the Ct shift only becomes obvious once you dilute away the surplus molecules.

Template degradation interacts with amplicon design too. A molecular sexing assay developed for emperor penguins targeted short sequences within a conserved exon of the CHD1 gene specifically because commonly used primers produced larger amplicons that were difficult to amplify from degraded DNA, and the authors validated sensitivity through a cross-factorial fragmentation and dilution experiment on synthetic templates [3]. If your template comes from forensic, archival, or environmental material, expect the tolerable dilution factor to shrink as the amplicon grows.

## Serial Dilutions: When One Step Is Not Enough

A serial dilution is a chain of smaller dilutions, each made from the previous one. Three sequential 1:10 dilutions produce a 1:1000 final dilution, and each intermediate tube is itself a usable standard.

```python
def serial(c0, df, n_steps):
    return [c0 / (df ** i) for i in range(n_steps + 1)]

concentrations = serial(85.0, 10, 3)
for i, c in enumerate(concentrations):
    print(f"Tube {i}: {c:.3f} ng/uL")
```

Output:

```text
Tube 0: 85.000 ng/uL
Tube 1: 8.500 ng/uL
Tube 2: 0.850 ng/uL
Tube 3: 0.085 ng/uL
```

Serial dilutions earn their keep in three situations. First, when the total dilution factor is large, since a single 1:1000 step asks your pipette to deliver 0.1 µL into 100 µL, which no manual pipette does reliably. Second, when you want a standard curve, since qPCR quantification depends on a ladder of known concentrations spanning several logs. Third, when carryover inhibitors are a concern. The Relay-MSRE-qPCR method addresses buffer incompatibility and inhibitor carryover by transferring each heat-inactivated digest to a fresh tube, diluting it at least 5-fold to attenuate inhibitors, and then reconstituting the next enzyme's buffer to 1× concentration, with glycerol kept at or below 5 percent by volume per step [4]. The same staged logic, dilute, transfer, re-buffer, applies whenever a reaction must tolerate the leftovers of a previous one.

Serial dilutions also appear in fields far from routine PCR. Forensic validation studies seed substrates with neat and 1:1000 dilutions of blood, semen, and saliva to probe the sensitivity limits of DNA recovery methods, and profile the diluted extracts to see where short tandem repeat analysis fails [5]. DNA-encoded library chemistry operates under deliberately dilute, aqueous, DNA-compatible conditions, which constrains which reactions are feasible on a DNA tag [6]. In every case, the dilution factor is chosen relative to the sensitivity of the readout, not pulled from habit.

## DNA Dilution Calculator Tips

These tips separate a correct number from a correct experiment.

1. **Quantify before you dilute, every time.** A stock concentration from last month is not data. Freeze-thaw cycles, evaporation, and evaporation-driven condensation on tube walls all change effective concentration.

2. **Prefer two-step dilutions for factors above roughly 1:50.** Two comfortable pipetting steps beat one desperate one. Plan the intermediate concentration so both steps use reasonable volumes, for example 85 ng/µL to 8.5 ng/µL to 0.85 ng/µL rather than 85 to 0.085 in one stroke.

3. **Make enough volume.** Aim for a final volume you can actually pipette and remeasure, commonly at least 20 µL, so that the diluted stock survives pipetting dead volume and a repeat measurement.

4. **Match diluent to downstream chemistry.** Nuclease-free water for immediate use, low-EDTA TE for storage. If your template will pass through another enzymatic step before PCR, think about cumulative salt and glycerol loads the way the Relay-MSRE-qPCR workflow budgets them explicitly [4].

5. **Change tips between stock and diluent.** Carryover of concentrated stock into the diluent tube quietly shifts every subsequent dilution made from it.

6. **Log the dilution factor, not just the result.** A tube labeled "5 ng/µL" is unfalsifiable. A tube labeled "85 ng/µL stock, 1:17, water, 2026-11-11" can be checked, redone, or debugged.

7. **Use copy number, not mass, when the assay demands it.** Digital PCR platforms report absolute copy number, and one multi-platform evaluation found that absolute copy number concentrations varied substantially across 14 commercial systems, with relative standard deviations of platform means between 6.05 and 12.71 percent, while copy number ratios measured in a duplex assay were far more consistent, with relative standard deviations as low as 0.47 percent [7]. If you move templates between platforms, ratios and relative dilutions travel better than absolute values.

8. **Validate with spike and recovery when the matrix is messy.** When DNA is diluted into a complex background such as serum, one [circulating tumor DNA](/knowledge/molecular-biology/circulating-tumor-dna) detection system reported mean spike recoveries between 98.59 and 101.61 percent in diluted human serum over three days, with intra-day relative standard deviations of 1.01 to 3.57 percent [8]. Recoveries in that band indicate the dilution and matrix together are behaving; large deviations point to adsorption or inhibition.

## Common Mistakes and Limitations

**Using the Nanodrop's back-calculated value without checking purity.** The Nanodrop assumes your sample is clean dsDNA. Contaminating RNA or free nucleotides inflate the 260 nm signal, so your true template concentration may be lower than displayed, and every dilution built on that number under-delivers template. Check the 260/280 and 260/230 ratios and, for concentration-critical work, confirm with a fluorometric assay.

**Pipetting less than your pipette can handle.** A calculated stock volume of 0.4 µL is not a measurement, it is a hope. Redo the dilution in two steps so no single transfer falls at the bottom of the tip's range.

**Confusing dilution factor with fold concentration.** A "1:10 dilution" means one part stock plus nine parts diluent, a 10-fold reduction in concentration. Some protocols mean one part stock plus ten parts diluent, an 11-fold reduction. When the stakes are high, state the final concentration explicitly instead of the ratio.

**Assuming dilution fixes inhibition.** Dilution reduces inhibitor concentration alongside template concentration, so it can rescue an inhibited reaction, but it simultaneously reduces signal. The staged-dilution design in Relay-MSRE-qPCR works because each dilution step is paired with buffer reconstitution rather than relying on dilution alone [4].

**Forgetting that damaged DNA changes the rules.** Autoclaving profoundly reduced the amplifiability of bee pathogen DNA in beeswax, rendering Paenibacillus larvae targets undetectable with Cq values above 40, while Melissococcus plutonius DNA remained detectable in 5 of 12 previously positive wax samples [9]. Chemically intact-looking concentration readings can therefore coexist with templates that no longer amplify. A concentration number describes molecules present, not molecules amplifiable.

**Chasing precision the instrument cannot support.** Absorbance measurements at low nanogram concentrations are noisy, and digital PCR platforms disagree on absolute copy number by several percent to more than ten percent depending on the system and partition volume assumptions [7]. Reporting your diluted template as "5.0 ng/µL" implies more certainty than the chain of measurements behind it usually justifies.

**Ignoring minimum template requirements of the specific assay.** Assays differ widely in the template range over which they respond linearly. A biosensor that performs well from 10 to 10^7 copies per µL compresses its response at 10^8 copies per µL [1]. Read your assay's stated range and dilute into the middle of it, not just below the ceiling.

**Blaming the calculator for experimental failures.** The arithmetic of C1V1 = C2V2 has no failure modes. When a dilution misbehaves, the causes are upstream (bad quantification, wrong units, degraded stock) or downstream (inhibition, poor amplicon design, pipetting error). Debug in that order.

One overall limitation deserves plain language: a calculator produces a plan, and only a re-measurement or an amplification result confirms that the plan worked. Individual samples, especially degraded or inhibitor-laden ones, may need empirical adjustment regardless of what the arithmetic says.

## How to Check That Your Dilution Worked

Three checks, in increasing cost.

First, re-quantify. Measure the diluted stock and compare it with the calculated target. A fluorometer is more informative than absorbance here because it responds to double-stranded DNA specifically. agreement within roughly 10 to 20 percent is realistic for dilute solutions.

Second, run a Ct comparison. Amplify both the stock and the diluted template with the same qPCR assay. If the dilution factor was 10-fold, the Ct should shift by about 3.3 cycles. A much larger shift suggests that some of the template is not amplifiable, a pattern consistent with damage that only becomes visible once surplus copies are diluted away [2].

Third, check linearity across a small dilution series. Two or three points should fall on a line with near-ideal cycle spacing. Curvature or wild points point to inhibition, and adding a step of dilution or clean-up, as staged-dilution workflows do [4], is the standard first response.

## Frequently Asked Questions

### How do I calculate a DNA dilution without a calculator?

Use C1 × V1 = C2 × V2 and solve for the unknown. Divide the target concentration by the stock concentration to get the fraction of the final volume that must be stock, multiply by your final volume, and add diluent to reach the total. A 17-fold dilution of a 100 µL working solution needs 5.88 µL of stock and 94.12 µL of diluent.

### Can I use Nanodrop readings directly for PCR template dilution?

Yes, but with caution. Nanodrop readings are fast and reagent-free, yet they overestimate amplifiable DNA when RNA, free nucleotides, or contaminants absorb at 260 nm, and they lose accuracy at low concentrations. For concentration-critical work, confirm with a dsDNA-specific fluorescent assay before diluting.

### How much DNA do I put in a PCR reaction?

Follow your protocol or polymerase manufacturer's recommendation, since the right amount depends on amplicon length, reaction volume, and assay type. As a scale reference, one published biosensor assay responded linearly from about 10 to 10^7 template copies per µL, and a 5 ng/µL stock of a 500 bp fragment contains roughly 9 billion copies per µL, so most routine reactions operate far below template saturation.

### What is a serial dilution and when should I use one?

A serial dilution is a chain of repeated smaller dilutions, such as three consecutive 1:10 steps to reach 1:1000. Use one whenever the total dilution factor is too large for accurate single-step pipetting, whenever you need a standard curve of several known concentrations, or whenever you want intermediate concentrations saved as usable stocks.

### Why does my diluted DNA amplify worse than expected?

The usual causes are inhibitors carried over from extraction, template damage, or an inaccurate starting concentration. Dilution alone can relieve inhibition, but staged workflows that pair each dilution with fresh buffer and a tube change handle it more reliably. If Ct shifts exceed the expected 3.3 cycles per 10-fold dilution, suspect that a fraction of the template is not amplifiable.

### How accurate are Nanodrop measurements at low DNA concentration?

Not very. Absorbance-based readings become unreliable in the low nanogram per microliter range because the signal is small and contaminants contribute proportionally more. Fluorometric assays with dsDNA-specific dyes are the better choice for quantifying dilute working stocks before or after a dilution.

### How do I convert ng/µL to copy number?

Divide the mass concentration in grams per microliter by the mass of one molecule, which is 660 g/mol per base pair times the fragment length, then multiply by Avogadro's number. For 5 ng/µL of a 500 bp double-stranded fragment, the result is about 9.1 × 10^9 copies per µL.

### Should I dilute DNA in water or TE buffer?

Use nuclease-free water for templates you will use immediately, and low-EDTA TE for stocks that will be stored or repeatedly freeze-thawed, because even small amounts of EDTA protect DNA from nuclease damage without meaningfully inhibiting polymerases at low concentrations. Avoid concentrated TE for PCR templates, since its EDTA can chelate the magnesium the reaction needs.

<script type="application/ld+json">
{
  "@context": "https://schema.org",
  "@type": "FAQPage",
  "mainEntity": [
    {
      "@type": "Question",
      "name": "How do I calculate a DNA dilution without a calculator?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Use C1 x V1 = C2 x V2 and solve for the unknown. Divide the target concentration by the stock concentration, multiply by the final volume to get the stock volume, and add diluent to reach the total. A 17-fold dilution to 100 uL needs 5.88 uL of stock and 94.12 uL of diluent."
      }
    },
    {
      "@type": "Question",
      "name": "Can I use Nanodrop readings directly for PCR template dilution?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Yes, with caution. Nanodrop readings can overestimate amplifiable DNA when RNA or contaminants absorb at 260 nm and lose accuracy at low concentrations. For concentration-critical work, confirm with a dsDNA-specific fluorescent assay before diluting."
      }
    },
    {
      "@type": "Question",
      "name": "How much DNA do I put in a PCR reaction?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Follow your protocol or polymerase manufacturer's recommendation, since the right amount depends on amplicon length, reaction volume, and assay type. As a reference, one published biosensor assay responded linearly from about 10 to 10^7 template copies per uL."
      }
    },
    {
      "@type": "Question",
      "name": "What is a serial dilution and when should I use one?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "A serial dilution is a chain of repeated smaller dilutions, such as three consecutive 1:10 steps to reach 1:1000. Use one when the total dilution factor is too large for accurate single-step pipetting, when building a standard curve, or when you want intermediate concentrations saved as stocks."
      }
    },
    {
      "@type": "Question",
      "name": "Why does my diluted DNA amplify worse than expected?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "The usual causes are inhibitors carried over from extraction, template damage, or an inaccurate starting concentration. Dilution can relieve inhibition, but staged workflows that pair dilution with fresh buffer handle it more reliably. Ct shifts beyond the expected 3.3 cycles per 10-fold dilution suggest some template is not amplifiable."
      }
    },
    {
      "@type": "Question",
      "name": "How accurate are Nanodrop measurements at low DNA concentration?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Not very. Absorbance-based readings become unreliable in the low nanogram per microliter range because the signal is small and contaminants contribute proportionally more. Fluorometric assays with dsDNA-specific dyes are the better choice for dilute working stocks."
      }
    },
    {
      "@type": "Question",
      "name": "How do I convert ng/uL to copy number?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Divide the mass concentration in grams per microliter by the mass of one molecule, which is 660 g/mol per base pair times the fragment length, then multiply by Avogadro's number. For 5 ng/uL of a 500 bp double-stranded fragment, the result is about 9.1 x 10^9 copies per uL."
      }
    },
    {
      "@type": "Question",
      "name": "Should I dilute DNA in water or TE buffer?",
      "acceptedAnswer": {
        "@type": "Answer",
        "text": "Use nuclease-free water for templates you will use immediately, and low-EDTA TE for stocks that will be stored or repeatedly freeze-thawed, since EDTA protects DNA from nuclease damage. Avoid concentrated TE for PCR templates because its EDTA can chelate the magnesium the reaction needs."
      }
    }
  ]
}
</script>

## Sources

1. [PCR-amplified blaKPC-2-triggered Cas12a activation coupled with DNA cube-confined Fc reporter cleavage on a Ti(3)C(2)/CsPbBr(3)@PDA@Au/GCE dual-mode ECL/SERS interface.](https://pubmed.ncbi.nlm.nih.gov/42986062/)
2. [Vacuum Drying Maintains the Integrity of DNA and Protects the Stability of PCR Amplification under Accelerated Aging.](https://pubmed.ncbi.nlm.nih.gov/42992133/)
3. [A Robust Assay for Molecular Sexing of Emperor Penguins Using Degraded DNA and the Potential Applicability to Other Spheniscids.](https://pubmed.ncbi.nlm.nih.gov/42992330/)
4. [Relay-MSRE-qPCR: A sequential multi-buffer staged-dilution workflow for methylation-sensitive restriction enzyme qPCR.](https://pubmed.ncbi.nlm.nih.gov/42990954/)
5. [The comparison of excision, swabbing, tape-lifting, and scraping as techniques for DNA evidence recovery: A preliminary study.](https://pubmed.ncbi.nlm.nih.gov/42835599/)
6. [On-DNA photochemical methods for expanding DNA-encoded library (DEL) chemical space.](https://pubmed.ncbi.nlm.nih.gov/42846366/)
7. [A multi-platform evaluation of digital PCR systems for DNA quantification: Implications for method standardization and metrological comparability.](https://pubmed.ncbi.nlm.nih.gov/43003564/)
8. [Construction of Y-shaped multivalent DNA walker system for sensitive detection of circulating tumor DNA.](https://pubmed.ncbi.nlm.nih.gov/43002260/)
9. [Detectability of Melissococcus plutonius and Paenibacillus larvae DNA in beeswax following thermal processing.](https://pubmed.ncbi.nlm.nih.gov/42957156/)