Pipetting Protein Solutions Without Foaming or Denaturation: A Guide

By Dr. Zubair Khalid, DVM, MS, PhD ·

Pipetting Protein Solutions Without Foaming or Denaturation: A Guide

Gentle pipetting of protein solutions preserves the molecule you worked weeks to purify. Done carelessly, a routine transfer can whip air into the sample, create a persistent foam cap, and expose the protein to an air-water interface where folded structures partially unfold and aggregate. Done well, the same transfer takes only a few extra seconds per tube. Hands-on time for a single transfer is under one minute once the technique is habit, and elapsed time for a typical batch of ten to twenty aliquots is five to ten minutes. There is no incubation or special equipment. The only requirements are a calibrated pipette, the right tips, and deliberate plunger control.

This guide explains why proteins foam and denature during liquid handling, walks through pipetting protein solutions without foaming step by step, and covers the pipetting technique details, pipette calibration, controls, troubleshooting, and the common mistakes that quietly ruin activity assays, crystallography drops, and NMR samples.

Why Protein Solutions Foam, and Why Foam Matters

Proteins are amphiphilic molecules. Part of each folded chain is hydrophobic and part is hydrophilic, so proteins naturally migrate to any air-water interface and adsorb there. Once adsorbed, a protein can partially unfold, spreading its hydrophobic regions into the air phase and its polar regions into the water. Repeated adsorption events build a viscoelastic film around each bubble, and that film is what makes protein foam persist long after the bubbles of a detergent-free buffer would have collapsed.

Research on protein foam films shows how substantial this interfacial behavior is. Studies of micellar sodium caseinate solutions using interferometric imaging revealed that protein foam films drain and stratify in structured steps, forming nanometer-scale ridges and flat domains at the moving thinning front, behavior that confirms proteins assemble into organized, persistent layers at the air-water interface [1]. In practical terms, a foam cap on your tube is not a cosmetic problem. Every bubble surface is a two-dimensional reactor where protein unfolds.

How much a given protein foams depends on the protein itself. Foaming capacity and foaming stability are recognized functional properties of protein preparations, and food protein research measures them routinely. Sodium caseinate is used commercially as a foaming agent precisely because it forms stable foams [1]. Mushroom-enriched milk protein preparations have been characterized specifically for their emulsifying and foaming properties [2], and silkworm pupae protein isolate showed foaming stability exceeding that of bovine serum albumin in a comparative extraction study [3]. For the bench scientist, the takeaway is that foaming tendency differs by protein, buffer, and concentration, so a liquid-handling routine that works for one purified protein may whip another into a froth.

Foam harms samples in three ways:

  1. Interfacial denaturation. Proteins adsorbed at bubble surfaces can unfold irreversibly. When bubbles collapse, the unfolded protein either refolds incorrectly or aggregates with neighbors.
  2. Loss of material. Foamy protein partitions into the foam layer, which clings to tube walls and tips and often never makes it into the recipient tube. For scarce samples, this is real yield loss.
  3. Measurement error. Foam in a cuvette or a microplate well scatters light and distorts absorbance readings, and foam in an NMR or crystallography sample introduces particulates and bubbles that ruin data collection.

Denaturation during pipetting has a second, quieter cause. Forcing liquid through the narrow orifice of a standard tip at high speed generates shear stress and local turbulence. Most soluble globular proteins tolerate moderate shear, but large multi-subunit complexes, membrane protein preparations, filamentous proteins, and fragile assemblies can lose activity from repeated high-shear transfers. Minimizing both interface exposure and shear is the goal of every tip and technique choice below.

The Principle of a Gentle Pipetting Technique

A foaming-free pipetting technique rests on five ideas:

  • Keep air out of the liquid. Aspirate with the tip submerged, release the plunger slowly, and never draw a bubble on purpose.
  • Minimize gas-liquid contact. Dispense onto the side wall of the receiving tube or below the liquid surface, so the stream does not plunge through the air into a pool.
  • Reduce shear. Use slow plunger speeds, wide-bore tips for fragile samples, and the smallest number of transfers possible.
  • Keep the protein off plastic surfaces it can denature on. Pre-wet tips, use low-retention or low-bind consumables, and avoid blow-out, which smears the last protein-rich film across the tip orifice.
  • Mix without whipping. Never vortex a protein solution. Mix by gentle inversion, slow rolling, or a few slow pipette strokes with a wide-bore tip.

Detergent choice matters when you formulate the buffer the protein will live in. Non-denaturing detergents such as sodium cholate preserve proteins in their native state, whereas harsher detergents such as SDS disrupt protein structure, a distinction demonstrated clearly in a tissue-clearing method that replaced SDS with sodium cholate specifically to retain native protein [4]. If your protein tolerates a mild, non-denaturing detergent, a small amount can also lower surface tension modestly and reduce foaming. Always test detergent compatibility with your assay first.

Materials and Equipment

ItemSpecification or working concentrationPurpose
Air-displacement pipettes (P2, P20, P100, P1000)Recently calibrated, clean, no leaking sealsAccurate, gentle liquid transfer
Positive-displacement pipetteOptional, for viscous samplesAccurate dispensing of glycerol-rich or concentrated solutions
Low-retention polypropylene tipsCertified low-retention, sterileLess protein and liquid clinging to tip walls
Wide-bore tipsStandard tips with the last few millimeters of the narrow end removed with sterile scissorsLower shear for fragile complexes
Protein stock solutionTypical bench range 0.1 to 5 mg/mL in a compatible bufferThe sample itself
Standard protein bufferFor example 50 mM Tris-HCl, pH 7.5, 150 mM NaCl, 5 percent glycerol, 1 mM DTT (adjust to your protein)Maintains pH, ionic strength, and stability
BSA practice solution1 mg/mL in the same bufferA safe, cheap stand-in for technique training
Low-bind microcentrifuge tubes0.5 or 1.5 mL, protein low-bindReduces adsorption losses
Benchtop mini centrifugeWith tube adaptersClears bubbles after handling
Ice bucket and pre-chilled tubesWet ice at 4 degrees CelsiusSlows proteolysis and aggregation
UV-Vis spectrophotometer or equivalent microvolume instrumentCapable of 2 µL measurementsRecovery checks by A280 or BCA assay

Safety note. Reducing agents such as DTT and beta-mercaptoethanol are irritants and, in the case of beta-mercaptoethanol, toxic by inhalation. Sodium azide, a common preservative in stock buffers, is acutely toxic. SDS, used in many protein workups, is an irritant [4][5]. Wear a lab coat, nitrile gloves, and eye protection, and handle all protein stocks under your institution's biosafety rules, especially recombinant or human-derived proteins.

Step-by-Step Guide: Pipetting Protein Solutions Without Foaming

Follow these steps in order. Each critical step includes the reason behind it, because knowing why a motion matters is what makes the technique stick.

  1. Check pipette calibration and tip fit before touching the sample. A pipette that delivers an inaccurate volume is a problem on its own, and a wobbly or ill-fitting tip leaks air into the aspiration stroke, which is a direct foam source. Confirm the tip clicks on squarely with even pressure around the shaft.
  1. Bring the sample and receiving tube to a consistent temperature and keep them on ice if the protein is cold-sensitive. Cold stock is more viscous, which changes pipetting dynamics, so note the temperature in your lab notebook. If you pipette cold, viscous solutions regularly, verify volume delivery with a gravimetric check rather than trusting the dial setting, because air-displacement pipettes are calibrated with water at room temperature.
  1. Pre-wet the tip. Aspirate a full volume of your buffer, dispense it to waste, then aspirate the protein solution. Reason: a dry polypropylene surface both attracts protein adsorption and promotes droplet clinging, and an unwetted tip's residual air film can seed bubbles. Pre-wetting equilibrates the humidity and surface chemistry inside the tip so the first delivery is accurate.
  1. Open the stock tube and, if the liquid meniscus is domed or sticky, briefly spin the tube in a mini centrifuge. Reason: liquid clinging to the cap and wall gets aspirated unevenly and drags air into the tip.
  1. Insert the tip a few millimeters below the liquid surface, off-center if the tube is small. Reason: a deeply submerged tip picks up back-pressure that drives fast, uncontrolled aspiration when the plunger releases, and a barely-submerged tip sucks air and surface foam. A few millimeters of depth gives a steady, bubble-free intake.
  1. Release the plunger slowly and smoothly to the first stop, then pause one to two seconds before withdrawal. Reason: fast plunger release is the single largest cause of foaming during aspiration. The pause ensures the full set volume has entered the tip before you move it, preventing partial volumes and air slugs.
  1. Withdraw the tip along the tube wall if possible, wiping the tip's outside only if liquid has climbed it. Reason: protein solution on the outside of the tip forms a droplet that later falls, splashes, and foams in the receiving tube.
  1. Angle the receiving tube and touch the tip to the wall, just above the liquid line if the tube already contains liquid, or against the lower wall if it is empty. Reason: dispensing down a wetted wall lets the protein sheet smoothly onto the surface instead of punching an air-plunging stream into the pool, which is the classic foaming event.
  1. Dispense to the first stop slowly, then pause before pushing to the second stop. Reason: the first stop delivers the calibrated volume. Pushing straight through to the second stop blows out residual liquid with force, aerosolizing and foaming the last of your protein.
  1. Decide deliberately whether to blow out. For most protein solutions, do not blow out. Leave the residual film in the tip. Reason: the residual film is protein-enriched and the blow-out stroke is where shear and bubbles peak. If your volumes are large enough that the residual matters, use reverse pipetting instead (next step).
  1. Use reverse pipetting for viscous solutions, very small volumes, or foamy stocks. Press the plunger past the first stop to the second stop, aspirate the set volume, and dispense only to the first stop with the tip against the wall. Discard the excess in the tip. Reason: reverse pipetting keeps positive pressure inside the tip throughout the stroke, which prevents dribbling and foaming at the orifice, at the cost of a small excess drawn into the tip. Check your pipette's documentation for the recommended extra volume, which manufacturers specify for reverse mode.
  1. Mix by inversion or slow pipette strokes, never by vortexing. For gentle mixing, invert the capped tube ten to twenty times, roll it between your palms, or draw and dispense three to five times with a wide-bore tip at slow speed. Reason: vortexing shear and air entrainment are exactly the conditions that denature interfacial protein and whip foam.
  1. If bubbles appear anyway, give the tube a short, gentle spin in a mini centrifuge. Reason: a low-speed pulse coalesces bubbles and drives them out of the liquid column without mechanical agitation. Do not flick or tap tubes containing precious samples, because flicking splashes protein onto the cap where it denatures and dries.
  1. Cap promptly, label, and return the stock to storage. Reason: every additional minute at room temperature, open to air, is additional interface time and evaporation, which raises concentration unpredictably in small volumes.

Pipette Calibration: The Foundation of Reliable Protein Handling

A gentle pipetting technique cannot rescue an inaccurate pipette. Pipette calibration is the periodic verification and adjustment of delivered volume against a traceable standard, typically by gravimetric testing in which the pipette dispenses water repeatedly onto an analytical balance and the delivered mass is converted to volume. Under the international standard governing piston-operated volumetric instruments, calibration accounts for test conditions including water temperature and evaporation, and it verifies performance at multiple points across the volume range, commonly near the top, middle, and lower limit of the pipette's span.

Why calibration matters specifically for protein work:

  • Small volumes amplify error. When you aliquot 5 µL of a scarce enzyme preparation, a pipette delivering 4.5 µL is a ten percent error in both concentration and downstream stoichiometry. Modern microvolume workflows push this further. A recently published ultra-sensitive protein quantification workflow requires only 2 µL of sample and detects concentrations as low as 0.01 mg/mL, and its improved reproducibility at low inputs underscores how much accuracy matters at microliter scale [6].
  • Viscosity and temperature skew air-displacement pipettes. Calibration is performed with water at room temperature. Cold, viscous protein solutions in high-salt or glycerol-containing buffers dispense differently, so the practical fix is reverse pipetting, positive-displacement pipettes, and periodic in-house verification.
  • Calibration drifts with use and abuse. Corroded pistons, worn seals, and dropped pipettes change delivery. Many institutional core facilities schedule calibration on a fixed cycle and spot-check frequently used pipettes gravimetrically between cycles. Follow your institution's schedule and the pipette manufacturer's documentation for service intervals.

A simple in-house check you can run in ten minutes: set the pipette to its nominal maximum, dispense ten replicate volumes of deionized water onto a tared analytical balance, and compute the mean delivered mass. Water at room temperature has a density near 1.0 mg per microliter, so milligrams read on the balance approximate microliters delivered. If the mean deviates substantially from the set volume, or the replicate spread is wide, send the pipette for professional calibration before trusting it with precious protein.

Controls and Verification of Your Technique

Validate your gentle pipetting technique the way you would validate any method, with comparisons:

  • Visual foam control. Pipette buffer alone and protein solution side by side using your technique, then photograph both tubes thirty seconds later. A trained technique leaves the protein tube clear with, at most, a few bubbles that dissipate.
  • Recovery control. Prepare a protein solution of known concentration, aliquot it with your technique into several tubes, and measure concentration in each aliquot by absorbance at 280 nanometers or a colorimetric assay such as BCA. Consistent readings across aliquots indicate that foam loss and adsorption are negligible. For scarce samples, low-volume quantification methods developed for 2 µL inputs make this check possible without spending material you cannot spare [6].
  • Activity control. For enzymes, run an activity assay on the stock and on freshly pipetted aliquots. A drop in specific activity after handling indicates interfacial or shear damage during transfer.
  • Practice control. Train with a 1 mg/mL BSA solution before handling a precious sample. BSA foams noticeably when mishandled, which gives you immediate visual feedback on your plunger speed.

Expected Results and How to Read Them

A correctly executed transfer produces a clear, homogeneous protein solution with no visible foam cap, no bubbles clinging to the tube walls, and no droplets inside the tip after dispensing. Concentration measurements across replicate aliquots should agree closely, and activity measurements should match the pre-handling stock within assay variability.

Read deviations like this:

  • A persistent foam layer means the air-water interface was disturbed. Look at your aspiration speed and whether you dispensed into the middle of the liquid pool rather than down the wall.
  • A hazy or particulate solution after handling suggests aggregation, which can follow interfacial denaturation or a buffer mismatch. Check pH, salt, and whether the protein prefers a mild non-denaturing additive [4].
  • Aliquot concentrations that drift downward across a series indicate adsorption or evaporation. Pre-wet tips, use low-bind tubes, work quickly, and cap tubes between transfers.
  • Loss of enzymatic activity with intact concentration points to shear or interface damage rather than material loss. Switch to wide-bore tips, slow the plunger further, and reduce the number of transfers.

Troubleshooting

SymptomLikely causeFix
Foam cap forms during aspirationPlunger released too fast, or tip too shallow and gulping airSlow the release, keep the tip a few millimeters under the surface, pre-wet the tip
Foam appears during dispensingStream dispensed into open liquid from height, or blow-out usedDispense down the angled tube wall, skip blow-out, use reverse pipetting
Persistent bubbles that will not popProtein forms a stable elastic film at bubble surfacesBrief low-speed spin in a mini centrifuge, touch bubbles with a clean hydrophobic pipette tip corner, add a compatible mild detergent after testing
Droplets cling to the tip exteriorLiquid climbed the outside of the tipWithdraw along the tube wall, wipe the exterior on the tube wall before moving, angle the receiving tube
Low recovery after aliquotingProtein adsorbed to tip and tube surfaces, or partitioned into foamUse low-retention tips and low-bind tubes, pre-wet tips with buffer containing salt, minimize transfers, avoid blow-out
Volumes inconsistent with viscous samplesAir-displacement pipette inaccurate in viscous liquidSwitch to reverse pipetting or a positive-displacement pipette, verify delivery gravimetrically
Enzyme loses activity after transferInterfacial or shear denaturationUse wide-bore tips, slow every plunger motion, reduce transfer count, keep the sample cold
Foam ruins a microplate readingBubbles in wells after dispensingDispense against the well wall near the liquid surface, tap-plate gently only if your assay tolerates it, or brief plate spin if your instrument supports it

Variations and Special Situations

Large-volume transfers. For milliliter-scale transfers with serological pipettes, set the pipette aid to low speed, keep the tip against the receiving vessel wall, and avoid letting the last drop fall from height. Low dispensing speed matters more than it does with micropipettes because the larger orifice moves more liquid per second.

Plate-based and automated workflows. In high-throughput protein workflows, dispensing mechanics remain experimental variables even under automation. A recent design-of-experiments study of transient protein production in CHO cells found that the pipetting sequence itself governed the outcome of a pre-complexing step, and that automated liquid handling with optimized settings roughly doubled volumetric productivity while cutting DNA input by about seventy percent, all without loss of product quality [7]. The lesson for manual pipettors is the same as for robots: the order, speed, and contact points of every transfer are part of the experiment. Automated liquid dispensers also reduce consumable waste and variability in plate-format protein preparation workflows, such as a 96-well alkaline-SDS lysis and precipitation protocol that extracts protein from 96 microbial samples in roughly thirty minutes using a bench-top dispenser [5].

Very small volumes. When your sample exists in microliters, every transfer counts twice, once for accuracy and once for adsorption loss. Low-volume quantification workflows built around 2 µL inputs exist precisely because microscale protein work demands it [6]. At this scale, pre-wetting, low-retention tips, and reverse pipetting stop being optional refinements.

Concentrated stocks for binding and titration experiments. Concentrated protein or ligand stocks are hard to handle without foaming, viscous drag, and instability. An NMR titration protocol designed to avoid concentrated ligand stocks altogether, generating intermediate titration points by swapping between two endpoint samples, demonstrates a broader principle. When handling a concentrated solution is the risky step, redesign the experiment so the risky step disappears [8].

Protein crystals and delicate assemblies. Crystallographers face the sharpest version of the problem. Protein crystals are damaged by abrupt changes in their surrounding solution, and humidity-controlled methods have been developed specifically to compensate for damage induced when crystals encounter alien solutions during handling [9]. The same logic applies to every sensitive sample. Change the environment around the protein slowly and gently, never by forceful mixing.

Proteins with high foaming tendency. Some preparations foam the moment air meets them. Casein-based solutions are used industrially as foaming agents because of their strong interfacial film formation [1], and foaming capacity is a deliberately optimized functional property in protein ingredient research [3][2]. If your protein belongs to this family, work at lower concentrations during transfers where possible, keep tips wetted at all times, and consider a validated, non-denaturing antifoam or mild detergent after confirming compatibility with your downstream assay.

Storage and Stability Notes

Gentle pipetting preserves protein only as long as storage preserves it. Standard practice for purified proteins:

  • Short term (hours to a few days). Store on wet ice or at 4 degrees Celsius in a buffered, stabilizing solution. Keep containers capped to limit evaporation and interface formation.
  • Long term. Aliquot into single-use volumes in low-bind tubes and freeze. Match the freeze format to the protein, whether flash-freezing in liquid nitrogen or controlled-rate freezing, and avoid repeated freeze-thaw cycles, because each cycle adds interfaces and shear during thawing.
  • Thawing. Thaw aliquots on ice or in the palm of a gloved hand. Never thaw protein in a warm water bath with agitation, and never vortex to speed thawing.
  • Additives. Glycerol at roughly 10 percent protects many proteins during freezing and reduces viscosity-driven pipetting surprises, though it makes accurate pipetting harder, which is where reverse pipetting or positive displacement earns its keep. Reducing agents such as DTT protect disulfide-dependent folds but degrade in air, so prepare them fresh.
  • Foam after thawing. If a thawed aliquot shows foam, spin it gently before use. Foam on a stored sample signals that it was agitated before freezing, and the protein that entered the foam layer may already be partially denatured.

Common Mistakes and Limitations

These are the errors seen most often at the bench, listed with the specific correction.

  1. Vortexing protein solutions to mix them. This is the most common and most damaging habit. Vortexing combines maximum shear with maximum air entrainment. Use inversion, rolling, or slow wide-bore pipette strokes.
  2. Blowing out every transfer out of habit. Blow-out is designed for water and reagent transfer, not protein. The final stroke is where foam and shear peak.
  3. Aspiring at full plunger speed. Speed feels efficient. It costs you yield and activity. Slow, smooth plunger motion is the core of pipetting protein solutions without foaming.
  4. Using dry tips on the first draw. The unpre-wetted tip both under-delivers and seeds adsorption. Pre-wet every tip that will touch protein.
  5. Dispensing from a height into the center of a tube. A falling stream entrains air. Angle the tube and touch the tip to the wall.
  6. Treating foam as harmless. Foam is visible evidence of interfacial protein, and protein in foam is frequently denatured protein. Clear it gently, then change the technique that produced it.
  7. Pipetting with an uncalibrated pipette. In protein work, concentration is the currency of every experiment. Pipette calibration is not a formality, it is the difference between 2.0 mg/mL and 1.7 mg/mL on a label.
  8. Assuming all proteins behave alike. Foaming behavior and interfacial stability vary widely among proteins, as the differences among caseins, albumins, and plant protein isolates demonstrate [1][3][10]. Validate handling for each new protein.
  9. Repetitive in-and-out pipetting to mix. Every aspiration cycle multiplies interface exposure. Three gentle strokes do the mixing job of twenty fast ones.

Limitations. This guide covers manual and semi-automated liquid handling of protein solutions. Individual proteins have idiosyncratic sensitivities to temperature, shear, pH, and interfaces, and a fragile sample may require specialized handling such as positive-displacement tools, custom wide-bore glassware, or exclusion of air exposure entirely. Follow your protein's published purification and handling literature, and consult the material's original characterization papers for known fragilities.

Frequently Asked Questions

Does foaming actually denature proteins?

Yes. Foaming concentrates protein at air-water bubble surfaces, where adsorbed molecules partially unfold and can aggregate when bubbles collapse. The persistent, elastic quality of protein foam, studied in detail for milk caseinate films [1], exists precisely because proteins form structured layers at interfaces, and that structural change at the interface is a form of denaturation.

Which pipetting mode is better for protein solutions, forward or reverse?

Reverse pipetting is better for viscous, foamy, or very small protein volumes because it maintains positive pressure in the tip and prevents dribbling and air ingestion. Forward pipetting with slow, smooth plunger control and no blow-out works well for aqueous, low-viscosity protein solutions in the middle of the tip's volume range.

How slow is slow when aspirating protein solutions?

Slow means the plunger release takes one to two seconds for a full hand motion rather than a snap, and the tip pauses in the liquid briefly before withdrawal. There is no universal seconds threshold because it depends on volume and viscosity, but if you can hear the liquid enter the tip or see bubbles, you are going too fast.

Do I need special low-retention tips for protein work?

They are strongly recommended for scarce or sticky proteins. Low-retention tips reduce the protein film left on the tip wall, which improves recovery and reduces the residual that would otherwise be blown out or wasted. Standard tips are acceptable for robust, abundant proteins when used with pre-wetting and no blow-out.

How often should pipettes be calibrated?

Follow your institution's calibration schedule and the pipette manufacturer's documentation. Most core facilities place frequently used pipettes on a fixed recalibration cycle and verify them gravimetrically in between. Any pipette that has been dropped, leaked, or delivered visibly inconsistent volumes should be pulled from service and calibrated immediately.

Can I vortex a protein solution briefly?

No. Even a short vortex burst introduces shear and air entrainment, and interfacial denaturation begins the moment bubbles form. Mix by gentle inversion, slow rolling, or a few slow wide-bore pipette strokes instead.

What should I do if foam has already formed?

Stop agitating the tube, let it sit briefly, then give it a short, gentle spin in a mini centrifuge to coalesce and release the bubbles. Then identify and correct the technique error that caused the foam, because the foam will return with the next transfer otherwise.

How do I handle very small volumes of precious protein?

Use the smallest calibrated pipette that comfortably covers the volume, pre-wet low-retention tips, use reverse pipetting, dispense down the tube wall, and never blow out. For quantification of such samples, microvolume workflows requiring as little as 2 µL have been developed specifically to preserve sample integrity at low inputs [6].

Sources

  1. Ultrathin Micellar Foam Films of Sodium Caseinate Protein Solutions.
  2. Goat’s Skim Milk Enriched with Agrocybe aegerita (V. Brig.) Vizzini Mushroom Extract: Optimization, Physico-Chemical Characterization, and Evaluation of Techno-Functional, Biological and Antimicrobial Properties
  3. Silkworm pupae protein isolate as a sustainable functional ingredient: Extraction optimization and application in sponge cake
  4. A novel protein-preserving passive tissue clearing approach using sodium cholate and urea for whole-organ imaging
  5. Alkaline-SDS cell lysis of microbes with acetone protein precipitation for proteomic sample preparation in 96-well plate format
  6. Enhanced protein extraction and quantification protocol for microsamples: An ultra-sensitive workflow for low-volume, low-concentration total protein lysates
  7. Harnessing Design of Experiments and Automation Significantly Improves Transient Protein Production in Chinese Hamster Ovary Cells and Accelerates Drug Discovery.
  8. A simulation-guided ‘swapping’ protocol for NMR titrations to study protein-protein interactions
  9. Electronic Reprint Structural Biology and Crystallization Communications Humidity Control Can Compensate for the Damage Induced in Protein Crystals by Alien Solutions Structural Biology and Crystallization Communications Structural Biology and Crystallization Communications Humidity Control Can Comp
  10. Block protocol for conventional profiling to sensory characterize plant protein isolates